Evaluation of an Automated Workstation for Aliquoting Whole Blood from Primary Sample Tubes
The Journal of Applied Laboratory Medicine · 2025
DOI: 10.1093/jalm/jfaf035 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, 2025. Verfasst von Eric Zimmerman Zuckerman und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, United States. Hauptthemen laut MEDLINE: Blood Specimen Collection, Automation, Laboratory. Die Arbeit nennt 11 Literaturangaben.
Abstract (Originalsprache)
Background
We evaluated the Tecan Fluent Mix and Pierce Workstation as an automated platform to aliquot directly from whole blood collection tubes. The Fluent Mix and Pierce Workstation uses the Tube Rotator™, an on-deck rotisserie, to keep whole blood in suspension, and 8 Piercing Tips™ to cross the pierceable caps of collection tubes to aliquot samples.
Methods
We performed accuracy and precision testing by aliquoting residual whole blood specimens into an on-deck gravimetric scale. To test carryover, we aliquoted alternating blanks and high calibrators for several analytes and performed mass spectrometry on the blanks. We performed 168 runs over a 21-day period to test the robustness of the platform. We aliquoted and analyzed 2 batches of samples on both our clinical platform and the Mix and Pierce Workstation to check concordance.
Results
We observed that the Fluent Mix and Pierce Workstation can aliquot 200 µL with 0.5% inaccuracy and 0.9%CV and can aliquot 300 µL with 0.1% inaccuracy and 0.7%CV. We were unable to detect any analyte carryover using mass spectrometry. We observed no hardware failures during the instrument stress test. Clinical comparison runs indicated that the performance of aliquots made on the Fluent Mix and Pierce Workstation was equivalent to that of aliquots made by our current clinical method.
Conclusions
Overall, our evaluation showed that the Tecan Fluent Mix and Pierce Workstation is a viable platform for aliquoting whole blood for downstream work flows and can eliminate labor-intensive sample decapping and recapping in high-throughput settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, United States.
- Senior Applications, Tecan US, Morrisville, NC, United States.
ORCID-Profile: Eric Zimmerman Zuckerman
Zitieren
APA
Zimmerman Zuckerman, E., Dukek, B., Hunsberger, D., Faber, J., Scheitz, E., Ward, A., Erickson, S., Hartman, T., Yiglletu, M., Baumann, N., & Jannetto, P. (2025). Evaluation of an Automated Workstation for Aliquoting Whole Blood from Primary Sample Tubes. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfaf035
Vancouver
Zimmerman Zuckerman E, Dukek B, Hunsberger D, Faber J, Scheitz E, Ward A, et al. Evaluation of an Automated Workstation for Aliquoting Whole Blood from Primary Sample Tubes. The Journal of Applied Laboratory Medicine. 2025. doi:10.1093/jalm/jfaf035
BibTeX
@article{ZimmermanZuckerman2025,
author = {Zimmerman Zuckerman, Eric and Dukek, Brian and Hunsberger, Dee and Faber, Jennifer and Scheitz, Eric and Ward, Amanda and Erickson, Steven and Hartman, Tom and Yiglletu, Martha and Baumann, Nikola and Jannetto, Paul},
title = {Evaluation of an Automated Workstation for Aliquoting Whole Blood from Primary Sample Tubes},
journal = {The Journal of Applied Laboratory Medicine},
year = {2025},
doi = {10.1093/jalm/jfaf035}
}