Vergleichsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, 2025. Verfasst von Monisha Parikh und 4 weiteren Autor:innen. Beteiligte Einrichtung: Medical Affairs, BD Life Sciences, Franklin Lakes, NJ, United States. Hauptthemen laut MEDLINE: Blood Specimen Collection, Blood Chemical Analysis. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Nontraditional sites for blood collection, including retail pharmacies, can expand access to laboratory testing. Three studies assessed the BD® MiniDraw™ Capillary Blood Collection System (BD MiniDraw) along with the BD MiniDraw SST™ Capillary Blood Collection Tube (BD MiniDraw SST tube). This system allows trained healthcare workers to collect capillary blood without requiring phlebotomy experience, in accordance with state laws.

Methods

Studies 1 and 2 evaluated clinical equivalence for selected serum chemistry analytes in the BD MiniDraw compared with currently marketed capillary and venous comparators. Study 3 evaluated within-tube stability for selected analytes up to 48 h in BD MiniDraw SST tubes following storage (2°–8°C). Contrived specimens were prepared to cover medically relevant ranges. Testing was performed on one general chemistry core laboratory analyzer in each study. Clinical equivalence was demonstrated if the biases and 95% limits were within a predefined clinical acceptance limit at all medical decision levels and time points for stability.

Results

Clinical equivalence was demonstrated for all analytes for BD MiniDraw vs capillary and venous comparators except alanine aminotransferase and chloride for the venous comparator in study 1. Both showed clinical equivalence at the upper limit of the reference range; the confidence interval exceeded the clinical acceptance limit at the lower limit. Within-tube stability was demonstrated in BD MiniDraw SST tubes up to 48 h for all selected analytes.

Conclusions

The BD MiniDraw presents an option for blood collection in nontraditional settings that is clinically equivalent to conventional capillary and venous collection for the specified analytes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Specimen CollectionBlood Chemical AnalysisHumansCapillariesPhlebotomyVeins

Autor:innen und Institutionen

  • Medical Affairs, BD Life Sciences, Franklin Lakes, NJ, United States.
  • Laboratory Development, Babson Diagnostics, Inc, Austin, TX, United States.
  • Research and Development, Babson Diagnostics, Inc, Austin, TX, United States.
  • Clinical Affairs, Babson Diagnostics, Inc, Austin, TX, United States.

Zitieren

APA

Parikh, M., Wimmer, C., DiPasquale, C., Barr, R. L., & Jacobson, J. W. (2025). Evaluation of a Novel Capillary Blood Collection System for Blood Sampling in Nontraditional Settings as Compared with Currently Marketed Capillary and Venous Blood Collection Systems for Selected General Chemistry Analytes. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfaf005

Vancouver

Parikh M, Wimmer C, DiPasquale C, Barr RL, Jacobson JW. Evaluation of a Novel Capillary Blood Collection System for Blood Sampling in Nontraditional Settings as Compared with Currently Marketed Capillary and Venous Blood Collection Systems for Selected General Chemistry Analytes. The Journal of Applied Laboratory Medicine. 2025. doi:10.1093/jalm/jfaf005

BibTeX
@article{Parikh2025,
  author = {Parikh, Monisha and Wimmer, Cathy and DiPasquale, Christopher and Barr, Roy L and Jacobson, James W},
  title = {Evaluation of a Novel Capillary Blood Collection System for Blood Sampling in Nontraditional Settings as Compared with Currently Marketed Capillary and Venous Blood Collection Systems for Selected General Chemistry Analytes},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2025},
  doi = {10.1093/jalm/jfaf005}
}
Kostenlos starten →