Comparison of 2 Serum Free Light Chain Assays with Creatinine Normal and Abnormal Populations Demonstrates the Need for Standardization
The Journal of Applied Laboratory Medicine · 2024
DOI: 10.1093/jalm/jfae065 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, 2024. Verfasst von Mark Griffiths und 4 weiteren Autor:innen. Beteiligte Einrichtung: Diazyme Laboratories Inc., Poway, CA, United States. Hauptthemen laut MEDLINE: Creatinine. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Background
The objective of this study was to compare The Binding Site’s Freelite on Optilite and Diazyme’s Kappa/Lambda free light chains (K/L FLC) on Abbott Architect c8000 with healthy and renal insufficient populations and to evaluate their respective reference intervals for serum free light chains (sFLCs).
Methods
Two hundred sixty serum samples were measured for creatinine and sFLCs by both assays and a subset by immunofixation electrophoresis. Verification of manufacturer-defined reference intervals was assessed.
Results
Kappa free light chains (KFLC) showed excellent correlation of 0.998 R2 with a slope of 0.73. For Lambda free light chains (LFLC), an acceptable correlation of 0.953 R2 was found with a slope of 1.50 as well as a skewness-based difference with a −12.70 intercept. Healthy estimated glomerular filtration rate (eGFR) ≥60 reference interval verification of central 95% could not be confirmed for either Freelite or Diazyme although LFLC was much closer than KFLC for both assays with Freelite KFLC recovering only 37% of values within reference interval claims. The K/L FLC ratio did not meet 100% claim for both Freelite (91%) and Diazyme (95%) among those with eGFR ≥60. Samples with eGFR ≤59 had increasingly higher levels of KFLC and LFLC for both assays. When comparing worsening eGFR status, Freelite recovered increasingly higher ratios while Diazyme recovered increasingly lower ratios.
Conclusions
Healthy reference intervals could not be verified for either Freelite or Diazyme. Renal reference intervals for Freelite are currently warranted while they are not recommended for Diazyme. The differences between these 2 assays can be minimized by standardization efforts such as recalibration.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Diazyme Laboratories Inc., Poway, CA, United States.
- Division of Clinical Biochemistry, University Health Network, Toronto, Canada.
- Abbott Laboratories, Abbott Park, IL, United States.
- Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, Canada.
Zitieren
APA
Griffiths, M., Cheng, P. L., Wang, X. Y., Schneider, R., & Kulasingam, V. (2024). Comparison of 2 Serum Free Light Chain Assays with Creatinine Normal and Abnormal Populations Demonstrates the Need for Standardization. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfae065
Vancouver
Griffiths M, Cheng PL, Wang XY, Schneider R, Kulasingam V. Comparison of 2 Serum Free Light Chain Assays with Creatinine Normal and Abnormal Populations Demonstrates the Need for Standardization. The Journal of Applied Laboratory Medicine. 2024. doi:10.1093/jalm/jfae065
BibTeX
@article{Griffiths2024,
author = {Griffiths, Mark and Cheng, Pow Lee and Wang, Xiao Yan and Schneider, Randal and Kulasingam, Vathany},
title = {Comparison of 2 Serum Free Light Chain Assays with Creatinine Normal and Abnormal Populations Demonstrates the Need for Standardization},
journal = {The Journal of Applied Laboratory Medicine},
year = {2024},
doi = {10.1093/jalm/jfae065}
}
EUVIMED