EUVIMEDEUVIMED entdecken
Validierungsstudie

Analytical and Clinical Validation of a Non-Ristocetin Based VWF Assay on 2 Automated Analyzers in a Large Reference Laboratory

Vivek Kumar, Debra Goode, Laura A Worfolk, Jeanne Rhea-McManus, John V Mitsios, Edward C C Wong iD

The Journal of Applied Laboratory Medicine · 2024

DOI: 10.1093/jalm/jfae063 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, 2024. Verfasst von Vivek Kumar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Quest Diagnostics, Secaucus, NJ, United States. Hauptthemen laut MEDLINE: von Willebrand Factor, Ristocetin. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Historically, von Willebrand factor (VWF) activity assays utilized ristocetin despite limitations including poor limits of detection and high imprecision. Newer VWF activity assays such as the INNOVANCE® VWF Ac assay, however, do not rely on ristocetin to measure platelet-dependent VWF function. The purpose of this study was to evaluate the analytical and clinical performance of the Siemens Healthineers INNOVANCE VWF Ac Assay on the Siemens BCS® XP and the Sysmex® CS-2500 systems in a large reference laboratory setting.

Methods

Performance indicators for the INNOVANCE VWF Ac assay were the limit of quantitation (LoQ), precision, and method comparison. Method comparison studies were performed using remnant plasma patient samples from routine coagulation tests and analyzed using both the INNOVANCE VWF Ac assay and the Siemens Healthineers ristocetin-dependent BC von Willebrand Reagent.

Results

Evaluation of the INNOVANCE VWF Ac assay on the BCS® XP and CS-2500 systems demonstrated good precision and a lower LoQ compared to the BC von Willebrand Reagent. Method comparisons support the use of the INNOVANCE VWF Ac assay on the BCS® XP and CS-2500 systems to measure platelet-dependent VWF function. The INNOVANCE VWF Ac assay was able to further assist in von Willebrand disease classification in 6/7 (86%) samples when the result was below the LoQ for the BC von Willebrand Reagent (ristocetin cofactor activity).

Conclusions

These data are consistent with the 2021 American Society of Hematology/International Society on Thrombosis and Haemostasis/National Hemophilia Foundation/World Federation of Hemophilia von Willebrand disease guidelines that suggest using newer assays such as the INNOVANCE VWF Ac assay in place of ristocetin cofactor activity assays.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

von Willebrand FactorRistocetinHumansBlood Coagulation TestsReproducibility of Resultsvon Willebrand DiseasesAutomation, LaboratoryLimit of Detection

Substanzen

von Willebrand FactorRistocetin

Autor:innen und Institutionen

  • Quest Diagnostics, Secaucus, NJ, United States.
  • Siemens Healthcare Diagnostics Inc., Tarrytown, NY, United States.
  • Department of Pediatrics, George Washington School of Medicine and Health Sciences, Washington, DC, United States.
  • Department of Pathology, George Washington School of Medicine and Health Sciences, Washington, DC, United States.

ORCID-Profile: Edward C C Wong

Zitieren

APA

Kumar, V., Goode, D., Worfolk, L. A., Rhea-McManus, J., Mitsios, J. V., & Wong, E. C. C. (2024). Analytical and Clinical Validation of a Non-Ristocetin Based VWF Assay on 2 Automated Analyzers in a Large Reference Laboratory. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfae063

Vancouver

Kumar V, Goode D, Worfolk LA, Rhea-McManus J, Mitsios JV, Wong ECC. Analytical and Clinical Validation of a Non-Ristocetin Based VWF Assay on 2 Automated Analyzers in a Large Reference Laboratory. The Journal of Applied Laboratory Medicine. 2024. doi:10.1093/jalm/jfae063

BibTeX
@article{Kumar2024,
  author = {Kumar, Vivek and Goode, Debra and Worfolk, Laura A and Rhea-McManus, Jeanne and Mitsios, John V and Wong, Edward C C},
  title = {Analytical and Clinical Validation of a Non-Ristocetin Based VWF Assay on 2 Automated Analyzers in a Large Reference Laboratory},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2024},
  doi = {10.1093/jalm/jfae063}
}