Evaluation of a Rapid Drug Test Device for Urine Fentanyl Compared to Mass Spectrometry and 2 Urine Fentanyl Assays

Erving T Laryea, James H Nichols iD

The Journal of Applied Laboratory Medicine · 2024

DOI: 10.1093/jalm/jfae059 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, 2024. Verfasst von Erving T Laryea und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pathology, Microbiology and Immunology, Vanderbilt University Medical Center, Nashville, TN, United States. Hauptthemen laut MEDLINE: Fentanyl, Substance Abuse Detection, Mass Spectrometry. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

A new Rapid Drug Test Device (RDTD) is available for analysis of urine fentanyl. With the rise in fentanyl abuse in the United States, we evaluated the analytical performance of the RDTD test strip compared to mass spectrometry and 2 urine fentanyl immunoassays.

Methods

Leftover, deidentified urine samples collected from inpatients and outpatients from our psychiatric hospital and other clinics were frozen at <−70°C, thawed at room temperature, and centrifuged. Aliquots were tested with the RDTD (CLIA Waived, Inc.) test strips and 2 urine fentanyl immunoassays: the ARK Fentanyl II assay (ARK Diagnostics Inc.) and the Immunalysis SEFRIA Fentanyl assay (Immunalysis Corporation). Both assays were conducted on the Abbott Alinity c chemistry analyzer (Abbott Laboratories). Mass spectrometry analysis was performed at ARUP Laboratories. All assays had a 1 ng/mL positive cutoff.

Results

A total of 142 urine samples included 70 positive and 72 negative samples. The RDTD strips had lower sensitivity (84.3%) and efficiency (85.9%) and showed a specificity of 87.5% compared to the other assays. The ARK Fentanyl II assay showed identical sensitivity (95.7%) to the Immunalysis SEFRIA Fentanyl assay but had higher specificity (94.4% vs 81.9%) and overall efficiency (95.1% vs 88.7%).

Conclusions

Differences were noted in the number of false negatives and positives among the assays. The RDTD demonstrated acceptable performance in detecting urine fentanyl in our patient population and would provide faster test results at point-of-care testing sites in our healthcare enterprise.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

FentanylSubstance Abuse DetectionMass SpectrometryHumansImmunoassaySensitivity and SpecificityAnalgesics, OpioidReagent Strips

Substanzen

FentanylAnalgesics, OpioidReagent Strips

Autor:innen und Institutionen

  • Department of Pathology, Microbiology and Immunology, Vanderbilt University Medical Center, Nashville, TN, United States.

ORCID-Profile: James H Nichols

Zitieren

APA

Laryea, E. T., & Nichols, J. H. (2024). Evaluation of a Rapid Drug Test Device for Urine Fentanyl Compared to Mass Spectrometry and 2 Urine Fentanyl Assays. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfae059

Vancouver

Laryea ET, Nichols JH. Evaluation of a Rapid Drug Test Device for Urine Fentanyl Compared to Mass Spectrometry and 2 Urine Fentanyl Assays. The Journal of Applied Laboratory Medicine. 2024. doi:10.1093/jalm/jfae059

BibTeX
@article{Laryea2024,
  author = {Laryea, Erving T and Nichols, James H},
  title = {Evaluation of a Rapid Drug Test Device for Urine Fentanyl Compared to Mass Spectrometry and 2 Urine Fentanyl Assays},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2024},
  doi = {10.1093/jalm/jfae059}
}
Kostenlos starten →