A Mathematical Algorithm for Dried Blood Spot Quality Assessment and Results concerning Quality from a Newborn Screening Program
The Journal of Applied Laboratory Medicine · 2024
DOI: 10.1093/jalm/jfae003 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, 2024. Verfasst von Oana R Oprea und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Biochemistry and Immunology, "George Emil Palade" University of Medicine, Pharmacy, Science, and Technology, Targu Mures, Roma…. Hauptthemen laut MEDLINE: Neonatal Screening, Dried Blood Spot Testing, Algorithms. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 16 Literaturangaben.
Abstract (Originalsprache)
Background
In addition to newborn screening, dried blood spots (DBSs) are used for a wide variety of analytes for clinical, epidemiological, and research purposes. Guidelines on DBS collection, storage, and transport are available, but it is suggested that each laboratory should establish its own acceptance criteria.
Methods
An optical scanning device was developed to assess the quality of DBSs received in the newborn screening laboratory from 11 maternity wards between 2013 and 2018. The algorithm was adjusted to agree with the visual examination consensus of experienced laboratory personnel. Once validated, the algorithm was used to categorize DBS specimens as either proper or improper. Improper DBS specimens were further divided based on 4 types of specimen defects.
Results
In total, 27 301 DBSs were analyzed. Compared with an annual DBS rejection rate of about 1%, automated scanning rejected 26.96% of the specimens as having at least one defect. The most common specimen defect was multi-spotting (ragged DBS, 19.13%). Among maternity wards, improper specimen rates varied greatly between 5.70% and 49.92%.
Conclusions
Improper specimen rates, as well as the dominant type of defect(s), are mainly institution-dependent, with various maternity wards consistently showing specific patterns of both parameters over time. Although validated in agreement with experienced laboratory personnel consensus, automated analysis rejects significantly more specimens. While continuous staff training, specimen quality monitoring, and problem-reporting to maternities is recommended, a thorough quality assessment strategy should also be implemented by every newborn screening laboratory. An important role in this regard may be played by automation in the form of optical scanning devices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Clinical Biochemistry and Immunology, "George Emil Palade" University of Medicine, Pharmacy, Science, and Technology, Targu Mures, Romania.
- Department of Electrical Engineering and Informatics, "George Emil Palade" University of Medicine, Pharmacy, Science, and Technology, Targu Mures, Romania.
ORCID-Profile: Albert Z Barabas
Zitieren
APA
Oprea, O. R., Barabas, A. Z., Manescu, I. B., & Dobreanu, M. (2024). A Mathematical Algorithm for Dried Blood Spot Quality Assessment and Results concerning Quality from a Newborn Screening Program. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfae003
Vancouver
Oprea OR, Barabas AZ, Manescu IB, Dobreanu M. A Mathematical Algorithm for Dried Blood Spot Quality Assessment and Results concerning Quality from a Newborn Screening Program. The Journal of Applied Laboratory Medicine. 2024. doi:10.1093/jalm/jfae003
BibTeX
@article{Oprea2024,
author = {Oprea, Oana R and Barabas, Albert Z and Manescu, Ion B and Dobreanu, Minodora},
title = {A Mathematical Algorithm for Dried Blood Spot Quality Assessment and Results concerning Quality from a Newborn Screening Program},
journal = {The Journal of Applied Laboratory Medicine},
year = {2024},
doi = {10.1093/jalm/jfae003}
}
EUVIMED