Analytical Concordance of Total Vitamin D on a Fully Automated Random-Access LC–MS/MS Platform
The Journal of Applied Laboratory Medicine · 2023
DOI: 10.1093/jalm/jfad036 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, 2023. Verfasst von Putuma P Gqamana und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pathology and Laboratory Medicine, University of Rochester Medical Center, Rochester, NY, United States. Hauptthemen laut MEDLINE: Vitamin D, Tandem Mass Spectrometry. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 17 Literaturangaben.
Abstract (Originalsprache)
Background
The adoption of LC–MS/MS laboratory developed tests in the clinical laboratory is limited by many factors including the lack of automation. Recently, the Cascadion™ clinical analyzer was introduced as a fully automated random-access LC–MS/MS platform. Here, the analytical concordance between the platform and a Roche immunoassay analyzer was investigated for vitamin D analysis in human serum, including samples selected for high triglyceride levels.
Methods
Analytical precision was evaluated on 3 levels of QC samples (10, 30, and 90 ng/mL) within days (n = 4, 5 days) and between days (20 days). Assay comparison to the Roche was performed using reference samples from the CDC and CAP programs for accuracy. Concordance was also monitored using routine patient samples, as well as samples selected for elevated triglyceride levels (>250 mg/dL).
Results
Precision met manufacturer specifications (<10% CV and <15% bias), whereas the accuracy evaluations showed a linear fit (y = 0.97x − 1.1, r = 0.995) with 1:1 correlation to reference samples, independent of C-3-epi-vitamin D levels. A mean positive bias (11%) was observed for the Roche measurements in normal patient samples, whereas a mean negative bias (−8%) was observed in samples selected for elevated triglyceride levels.
Conclusions
Cascadion measurements of total vitamin D compared favorably with Roche results in our laboratory, although discordance was observed in the analysis of patient serum, which could be explained in terms of known differences between the 2 assays. However, operational issues need to be addressed to effect clinical adoption.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Pathology and Laboratory Medicine, University of Rochester Medical Center, Rochester, NY, United States.
ORCID-Profile: Putuma P Gqamana
Zitieren
APA
Gqamana, P. P., Militello, L., McMaster, J. M., Daley, S. J., & Zhang, Y. V. (2023). Analytical Concordance of Total Vitamin D on a Fully Automated Random-Access LC–MS/MS Platform. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfad036
Vancouver
Gqamana PP, Militello L, McMaster JM, Daley SJ, Zhang YV. Analytical Concordance of Total Vitamin D on a Fully Automated Random-Access LC–MS/MS Platform. The Journal of Applied Laboratory Medicine. 2023. doi:10.1093/jalm/jfad036
BibTeX
@article{Gqamana2023,
author = {Gqamana, Putuma P and Militello, Leah and McMaster, Jeffrey M and Daley, Stacy J and Zhang, Y Victoria},
title = {Analytical Concordance of Total Vitamin D on a Fully Automated Random-Access LC–MS/MS Platform},
journal = {The Journal of Applied Laboratory Medicine},
year = {2023},
doi = {10.1093/jalm/jfad036}
}
EUVIMED