Pacing Performance of a Modular Communicative Leadless Pacing-Defibrillator System
Europace · 05.11.2026
DOI: 10.1093/europace/euag309 ↗Kurz gefasst
Fachartikel in der Zeitschrift Europace, online erschienen am 05.11.2026. Verfasst von Lluís Mont und 23 weiteren Autor:innen. Beteiligte Einrichtung: Institut Clínic Cardiovascular, Hospital Clínic, Universitat de Barcelona. Barcelona, Spain. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background and Aims MODULAR ATP is an ongoing trial assessing the efficacy of a subcutaneous implantable cardioverter defibrillator (S-ICD) in communication with a leadless pacemaker (LP) that can deliver bradycardia pacing and antitachycardia pacing. The objective of this study was to assess safety and pacing performance of the investigational LP.
Methods
The study enrolled patients with an ICD indication and at high risk for monomorphic ventricular tachycardia. Pacemaker dependent patients were excluded. Primary safety endpoint 2 was the percentage of patients free from major LP-related complications at 12 months. The secondary effectiveness endpoint assessed the rate response performance of the LP under workload. Parameters related to the pacing performance were also measured.
Results
In 297 patients, the mean age was 59±12 years, the mean left ventricular ejection fraction was 35.3±13.1%, and 17.5% were female. LP implant was attempted in 286 patients and was successful in 100%. The major LP-related complication free rate at 12 months was 97.2% (one-sided 97.5% LCL 94.5%), surpassing the performance goal of 81%. There were no LP-related complications following Implant. For the secondary effectiveness endpoint, the slope of the normalized sensor rate versus the normalized workload was 0.96 (95% LCL 0.91, 95% UCL 1.02), within the performance goal range of 0.65-1.35. At 12 months, 96.8% of patients had a pacing capture threshold ≤ 2V @ 0.4ms pulse width.
Conclusions
The performance of the modular pacing-defibrillator system at one year demonstrated a low complication rate and reliable pacing functionality.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institut Clínic Cardiovascular, Hospital Clínic, Universitat de Barcelona. Barcelona, Spain.
- Emory University Section of Cardiac Electrophysiology, Atlanta, GA, USA.
- University Hospital Southampton, Southampton, United Kingdom.
- Institute John Schufeldt ASU School of Medicine and Medical Engineering and HonorHealth, Scottsdale, AZ, USA.
- Department of Cardiology, Liverpool Heart and Chest Hospital, Liverpool, United Kingdom.
- Department of Cardiovascular Medicine, Mayo Clinic, Rochester, MN, USA.
ORCID-Profile: Lluís Mont, Michael S Lloyd, Paul R Roberts, Rahul Doshi, David J Wright, Paul A Friedman, Petr Neuzil, Carina Blomstrom-Lundqvist, Maria Grazia Bongiorni, Martin Burke, Daniel Gras, Steven P Kutalek, Harish Manyam, Wilber Su, Gerardo Nigro, Tara Bachman, Reinoud E Knops, Vivek Y Reddy
Zitieren
APA
Mont, L., Lloyd, M. S., Roberts, P. R., Doshi, R., Wright, D. J., Friedman, P. A., Neuzil, P., Blomstrom-Lundqvist, C., Bongiorni, M. G., Burke, M., Gras, D., Kutalek, S. P., Plourde, B., Merchant, F. M., Manyam, H., Su, W., Nigro, G., Dunmire, S. K., Appl, U., …, & Reddy, V. Y. (2026). Pacing Performance of a Modular Communicative Leadless Pacing-Defibrillator System. Europace. https://doi.org/10.1093/europace/euag309
Vancouver
Mont L, Lloyd MS, Roberts PR, Doshi R, Wright DJ, Friedman PA, et al. Pacing Performance of a Modular Communicative Leadless Pacing-Defibrillator System. Europace. 2026. doi:10.1093/europace/euag309
BibTeX
@article{Mont2026,
author = {Mont, Lluís and Lloyd, Michael S and Roberts, Paul R and Doshi, Rahul and Wright, David J and Friedman, Paul A and Neuzil, Petr and Blomstrom-Lundqvist, Carina and Bongiorni, Maria Grazia and Burke, Martin and Gras, Daniel and Kutalek, Steven P and Plourde, Benoit and Merchant, Faisal M and Manyam, Harish and Su, Wilber and Nigro, Gerardo and Dunmire, Samantha K and Appl, Ursula and Carter, Nathan and Swackhamer, Bryan and Bachman, Tara and Knops, Reinoud E and Reddy, Vivek Y},
title = {Pacing Performance of a Modular Communicative Leadless Pacing-Defibrillator System},
journal = {Europace},
year = {2026},
doi = {10.1093/europace/euag309}
}
EUVIMED