Übersichtsarbeit in der Zeitschrift Europace, 2026. Verfasst von Ante Anić und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Disease, University Clinical Hospital, Split, Croatia. Hauptthemen laut MEDLINE: Irreversible Electroporation Therapy, Arrhythmias, Cardiac. Die Arbeit nennt 96 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Boston Scientific.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Pulsed field ablation (PFA) has emerged as a transformative approach for treating cardiac arrhythmias by exploiting irreversible electroporation to preferentially injure myocardium while limiting collateral damage. Translating this biophysical principle into safe and effective clinical systems requires careful integration of waveform design, catheter engineering, computational and experimental modelling, safety testing, and procedural workflow.

Objective

This review summarizes the fundamental physical and engineering principles that govern cardiac PFA and examines how these principles influence efficacy, safety, lesion formation, and clinical usability. Biophysical and technical considerations Electroporation is governed primarily by the local electric field and waveform-dependent lethal thresholds rather than generator voltage alone. Key waveform variables include monophasic vs. biphasic pulses, microsecond vs. nanosecond pulse widths, duty cycle, and total application duration. Catheter design further determines electric field distribution, lesion geometry, current density, and applicability across anatomic targets, with important distinctions between bipolar and monopolar configurations. Modelling and safety Experimental and computational approaches used to investigate PFA include finite element simulations, cell culture systems, vegetal substrates, and in vivo animal models, each with distinct strengths and limitations. High-voltage pulse delivery can also produce unintended effects, including Joule heating, electrochemical reactions, bubble formation, arcing, and interactions with implanted electronic devices. These phenomena highlight the importance of optimized waveform design, electrode configuration, and safety monitoring. Clinical usability Clinical adoption depends not only on biological efficacy and safety but also on workflow efficiency, catheter versatility, procedure duration, mapping integration, anaesthesia requirements, and adaptability to differing clinical practices and geographic settings.

Conclusions

The performance of a PFA system reflects the interaction of biophysics, waveform characteristics, catheter design, modelling, safety controls, and procedural usability. Understanding these interconnected principles is essential for developing PFA technologies that are effective, safe, predictable, and practical for contemporary electrophysiology laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Irreversible Electroporation TherapyArrhythmias, CardiacHumansAnimalsEquipment DesignTreatment OutcomeModels, CardiovascularCardiac CathetersComputer Simulation

Stichwörter der Autor:innen

Cardiac arrhythmiasCatheter designComputational modelingElectric fieldIrreversible electroporationPulsed field ablationWaveform design

Autor:innen und Institutionen

  • Department of Cardiovascular Disease, University Clinical Hospital, Split, Croatia.
  • Department of Engineering, Universitat Pompeu Fabra, Barcelona, Spain.
  • Department of Electrophysiology, Boston Scientific Corporation, 4100 Hamline Ave. N, St. Paul, MN 55112, USA.
  • Helmsley Electrophysiology Center, Mount Sinai Fuster Heart Hospital, New York, NY, USA.

ORCID-Profile: Ante Anić, Antoni Ivorra, Brendan Koop, Nathan J Kirkpatrick, Jacob Koruth

Förderung

  • Boston Scientific

Zitieren

APA

Anić, A., Ivorra, A., Koop, B., Kirkpatrick, N. J., & Koruth, J. (2026). Pulsed field ablation: from concepts to technology. Europace. https://doi.org/10.1093/europace/euag205

Vancouver

Anić A, Ivorra A, Koop B, Kirkpatrick NJ, Koruth J. Pulsed field ablation: from concepts to technology. Europace. 2026. doi:10.1093/europace/euag205

BibTeX
@article{Ani2026,
  author = {Anić, Ante and Ivorra, Antoni and Koop, Brendan and Kirkpatrick, Nathan J and Koruth, Jacob},
  title = {Pulsed field ablation: from concepts to technology},
  journal = {Europace},
  year = {2026},
  doi = {10.1093/europace/euag205}
}
Kostenlos starten →