Reprocessing of single-use medical devices in cardiology: a systematic literature review of safety and performance characteristics applied to cardiac electrophysiology
Europace · 2026
DOI: 10.1093/europace/euaf316 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Europace, 2026. Verfasst von Samantha Huynh und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmacy, European Georges Pompidou Hospital, AP-HP, Paris, France. Hauptthemen laut MEDLINE: Equipment Reuse, Electrophysiologic Techniques, Cardiac, Cardiac Catheters, Equipment Contamination. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aims
The rising burden of cardiovascular diseases, especially atrial fibrillation, has increased demand for electrophysiology (EP) procedures in Europe, driving greater reliance on disposable devices like diagnostic and ablation catheters. Reprocessing single-use medical devices offers potential economic and environmental benefits, yet concerns persist regarding device integrity and safety. Under the European Union Medical Device Regulation, reprocessing is permitted if national laws allow it; however, implementation varies across Member States. This systematic literature review evaluates the safety and performance of reprocessed cardiac EP catheters originally intended for single use. Methods and results Following PRISMA guidelines, literature searches of PubMed and Embase identified in vitro and in vivo studies that examined the safety and functionality of reprocessed EP catheters. Key outcomes included infection risk, device sterility, mechanical and electrical integrity, and adverse events. Twelve studies (four in vivo and eight in vitro) involving >1200 patients and multiple catheter brands were included. Under stringent reprocessing protocols, reprocessed EP catheters showed comparable safety and mechanical performance to new devices. However, gaps remain in evidence regarding prion and fungal contamination, the maximum number of safe reprocessing cycles, and the detection of rare complications.
Conclusion
The absence of adverse events and reliable performance associated with reprocessed EP catheters reported in this study may encourage European countries that have not yet authorized single-use medical device reprocessing to consider its adoption. However, the broader implementation of this practice remains contingent on its applicability and logistical feasibility within each national context.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Pharmacy, European Georges Pompidou Hospital, AP-HP, Paris, France.
- School of Pharmacy and Pharmaceutical Sciences, Trinity College Dublin, Dublin, Ireland.
- Groupe Parisien Universitaire de Rythmologie (GPUR/AP-HP), Paris, France.
- Department of Cardiology, Bichat Hospital, AP-HP, Paris, France.
- Paris-Cité University, Paris, France.
- Department of Cardiology, La Pitié Salpêtrière Hospital, AP-HP, Paris, France.
ORCID-Profile: Samantha Huynh, Fiona Murphy, Brigitte Sabatier, Fabrice Extramiana, Estelle Gandjbakhch, Nicolas Lellouche, Tess Martin, Eloi Marijon, Nicolas Martelli
Zitieren
APA
Huynh, S., Murphy, F., Sabatier, B., Pineau, J., Extramiana, F., Gandjbakhch, E., Lellouche, N., Martin, T., Marijon, E., & Martelli, N. (2026). Reprocessing of single-use medical devices in cardiology: a systematic literature review of safety and performance characteristics applied to cardiac electrophysiology. Europace. https://doi.org/10.1093/europace/euaf316
Vancouver
Huynh S, Murphy F, Sabatier B, Pineau J, Extramiana F, Gandjbakhch E, et al. Reprocessing of single-use medical devices in cardiology: a systematic literature review of safety and performance characteristics applied to cardiac electrophysiology. Europace. 2026. doi:10.1093/europace/euaf316
BibTeX
@article{Huynh2026,
author = {Huynh, Samantha and Murphy, Fiona and Sabatier, Brigitte and Pineau, Judith and Extramiana, Fabrice and Gandjbakhch, Estelle and Lellouche, Nicolas and Martin, Tess and Marijon, Eloi and Martelli, Nicolas},
title = {Reprocessing of single-use medical devices in cardiology: a systematic literature review of safety and performance characteristics applied to cardiac electrophysiology},
journal = {Europace},
year = {2026},
doi = {10.1093/europace/euaf316}
}
EUVIMED