Haemolysis and myocardial and neural injury after monopolar pulsed field ablation using a novel lattice-tip catheter to treat atrial fibrillation
Europace · 2025
DOI: 10.1093/europace/euaf210 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Europace, 2025. Verfasst von Christian Gold und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology and Vascular Medicine, University Heart and Vascular Center Frankfurt, Goethe University Frankfurt, Theodor-Stern Kai 7, Fra…. Hauptthemen laut MEDLINE: Atrial Fibrillation, Hemolysis, Catheter Ablation, Cardiac Catheters. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aims
The aim of this study was to assess the risk of haemolysis and the extent of myocardial and neural injury after monopolar, biphasic pulsed field ablation (PFA) using a lattice-tip catheter in comparison to single-shot PF ablation platforms employing bipolar, biphasic waveforms. Methods and results This prospective study included consecutive patients undergoing PFA for atrial fibrillation (AF) using the Affera™ mapping and ablation system (n = 40). Biomarkers for haemolysis (haptoglobin, lactate dehydrogenase, bilirubin), myocardial injury [high-sensitive troponin T, creatine kinase (CK), creatine kinase MB (CK-MB)], neurocardiac injury (S100), and renal function (creatinine) were assessed pre- and within 24 h post-ablation. A subgroup analysis of first-time pulmonary vein isolation-only procedures compared biomarker changes across Affera™, Farapulse™ (PFA-F), and PulseSelect™ (PFA-P). Post-procedural haemolysis occurred across all PFA platforms. The decrease in Δhaptoglobin was most pronounced in PFA-F [AfferaTM: (−) 13.8 ± 18.5 vs. PFA-P: (−) 36.8 ± 35.9 vs. PFA-F: (−) 60.7 ± 26.3 mg/dL, P = <0.001], without haemolysis-related complications. AfferaTM shows a trend towards a higher increase in myocardial injury markers (Δtroponin, 1537 [580] vs. 970 [1023] vs. 1051 [592] pg/mL, P = 0.180; ΔCK, 232 [168] vs. 153 [132] vs. 102 [144] U/L, P = 0.006; ΔCK-MB, 28.5 [15.3] vs. 14.6 [12.4] vs. 13.6 [10.5] U/L, P = 0.055, for Affera TM, PFA-P, and PFA-F, respectively). After ablation, S100 increased in PFA-P and PFA-F, but not in AfferaTM.
Conclusion
Post-procedural haemolysis after PFA for AF treatment is common and occurs across all PFA platforms. Pulsed field ablation using AfferaTM results in more myocardial injury than bipolar PFA systems with no indication of neural damage.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Cardiology and Vascular Medicine, University Heart and Vascular Center Frankfurt, Goethe University Frankfurt, Theodor-Stern Kai 7, Frankfurt am Main 60590, Germany.
- Department of Cardiology, Maastricht University Medical Center and Cardiovascular Research Institute Maastricht, Maastricht, The Netherlands.
- Department of Biomedical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
ORCID-Profile: Victoria Johnson, Jana Kupusovic, Julia W Erath-Honold, Dominik Linz, David Leistner, Laura Rottner
Zitieren
APA
Gold, C., Pratz, P., Falagkari, A., Johnson, V., Post, F., Roth, E., Kupusovic, J., Erath-Honold, J. W., Linz, D., Leistner, D., Wakili, R., & Rottner, L. (2025). Haemolysis and myocardial and neural injury after monopolar pulsed field ablation using a novel lattice-tip catheter to treat atrial fibrillation. Europace. https://doi.org/10.1093/europace/euaf210
Vancouver
Gold C, Pratz P, Falagkari A, Johnson V, Post F, Roth E, et al. Haemolysis and myocardial and neural injury after monopolar pulsed field ablation using a novel lattice-tip catheter to treat atrial fibrillation. Europace. 2025. doi:10.1093/europace/euaf210
BibTeX
@article{Gold2025,
author = {Gold, Christian and Pratz, Paul and Falagkari, Anastasia and Johnson, Victoria and Post, Florian and Roth, Esther and Kupusovic, Jana and Erath-Honold, Julia W and Linz, Dominik and Leistner, David and Wakili, Reza and Rottner, Laura},
title = {Haemolysis and myocardial and neural injury after monopolar pulsed field ablation using a novel lattice-tip catheter to treat atrial fibrillation},
journal = {Europace},
year = {2025},
doi = {10.1093/europace/euaf210}
}
EUVIMED