Thermal side effects during pulsed field ablation: an analysis using computer modelling
Europace · 2025
DOI: 10.1093/europace/euaf035 ↗Kurz gefasst
Fachartikel in der Zeitschrift Europace, 2025. Verfasst von Marcela Mercado-Montoya und 5 weiteren Autor:innen. Beteiligte Einrichtung: Bioengineering Department, IN SILICO STEM S.A.S, Calle 52 81-48, Medellín 050035, Colombia. Hauptthemen laut MEDLINE: Computer Simulation, Models, Cardiovascular, Esophagus, Catheter Ablation. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NHLBI NIH HHS, Spanish Ministerio de Ciencia e Innovación, Agencia Estatal de Investigación, Fondo Europeo de Desarrollo Regional, NIH HHS.
Abstract (Originalsprache)
Aims
Pulsed field ablation (PFA) is described as non-thermal, but data from oncology and cardiology show thermal effects occur. The specific waveform parameters influencing thermal energy development during PFA are unclear. The aim of this study is to numerically evaluate the thermal effects of PFA on myocardial and oesophageal tissue at various peak voltage conditions. Methods and results A three-dimensional computer model of the left atrium quantified thermal effects from PFA at peak voltages of 1, 1.5, and 2 kV. Energy was applied using a bipolar configuration with far-field and symmetry boundaries set as electrically insulating. A monophasic waveform with a 100 μs pulse width and a 1 s gap between pulses was applied for a total of 50 pulses, mimicking clinical conditions. Minimal temperature rise in the oesophagus was observed with 1 kV pulses (214.5 J). At 1.5 and 2 kV (570.3 and 1.23 kJ), temperatures reached 46.3°C and >62°C, respectively, after a single pulse train. These findings suggest that repeated applications could lead to even higher temperatures, especially if good tissue contact is obtained. These results align with data from other medical fields using pulsed field treatments.
Conclusion
Thermal effects from PFA depend on the total energy deposited, with peak voltage being a significant factor. Current commercially available PFA systems have the potential to induce collateral thermal injury with repeated applications of pulsed field energy. This highlights the need for careful monitoring and adjustment of PFA parameters in clinical settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Bioengineering Department, IN SILICO STEM S.A.S, Calle 52 81-48, Medellín 050035, Colombia.
- Department of Electrophysiology, Scripps Health, La Jolla, CA, USA.
- Department of Emergency Medicine, University of Texas Southwestern Medical Center, Dallas, TX, USA.
- BioMIT, Department of Electronic Engineering, Universitat Politècnica de València, València, Spain.
- Department of Electrical and Computer Engineering, University of Texas at Dallas, Dallas, USA.
ORCID-Profile: Marcela Mercado-Montoya, Tatiana Gomez-Bustamante, Steven R Mickelsen, Erik Kulstad, Ana González-Suárez, Lawrence J Overzet
Förderung
- NHLBI NIH HHS
- Spanish Ministerio de Ciencia e Innovación, Agencia Estatal de Investigación, Fondo Europeo de Desarrollo Regional
- NIH HHS
- National Heart, Lung, And Blood Institute of the National Institutes of Health
- National Institutes of Health
Zitieren
APA
Mercado-Montoya, M., Gomez-Bustamante, T., Mickelsen, S. R., Kulstad, E., González-Suárez, A., & Overzet, L. J. (2025). Thermal side effects during pulsed field ablation: an analysis using computer modelling. Europace. https://doi.org/10.1093/europace/euaf035
Vancouver
Mercado-Montoya M, Gomez-Bustamante T, Mickelsen SR, Kulstad E, González-Suárez A, Overzet LJ. Thermal side effects during pulsed field ablation: an analysis using computer modelling. Europace. 2025. doi:10.1093/europace/euaf035
BibTeX
@article{MercadoMontoya2025,
author = {Mercado-Montoya, Marcela and Gomez-Bustamante, Tatiana and Mickelsen, Steven R and Kulstad, Erik and González-Suárez, Ana and Overzet, Lawrence J},
title = {Thermal side effects during pulsed field ablation: an analysis using computer modelling},
journal = {Europace},
year = {2025},
doi = {10.1093/europace/euaf035}
}
EUVIMED