Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients
Europace · 2024
DOI: 10.1093/europace/euae170 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Europace, 2024. Verfasst von Jonas Wörmann und 25 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrophysiology, Heart Center at the University of Cologne, Kerpener Str. 62, 50937 Cologne, Germany. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Registries, Electric Power Supplies. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Aims
Subcutaneous implantable cardioverter-defibrillators (S-ICDs) have become established in preventing sudden cardiac death, with some advantages over transvenous defibrillator systems, including a lower incidence of lead failures. Despite technological advancements, S-ICD carriers may suffer from significant complications, such as premature battery depletion (PBD), that led to an advisory for nearly 40 000 patients. This multicentre study evaluated the incidence of PBD in a large set of S-ICD patients. Methods and results Data from patients implanted with S-ICD models A209 and A219 between October 2012 and July 2023 across nine centres in Europe and the USA were reviewed. Incidence and implications of PBD, defined as clinically observed sudden drop in battery longevity, were analysed and compared to PBD with the definition of battery depletion within 60 months. Prospectively collected clinical data were obtained retrospectively from medical records, device telemetry, and manufacturer reports. This registry is listed on ClinicalTrials.gov (NCT05713708). Of the 1112 S-ICD devices analysed, 547 (49.2%) were equipped with a potentially affected capacitor linked to PBD occurrence, currently under Food and Drug Administration advisory. The median follow-up time for all patients was 46 [inter-quartile range (IQR) 24–63] months. Clinically suspected PBD was observed in 159 (29.1%) of cases, with a median time to generator removal or replacement of 65 (IQR 55–72) months, indicative of significant deviations from expected battery lifespan. Manufacturer confirmation of PBD was made in 91.7% of devices returned for analysis. No cases of PBD were observed in devices that were not under advisory.
Conclusion
This manufacturer-independent analysis highlights a notable incidence of PBD in patients equipped with S-ICD models under advisory, and the rate of PBD in this study corresponds to the rate currently estimated by the manufacturer. To the best of our knowledge, this provides the largest contemporary peer-reviewed study cohort investigating the actual incidence of PBD in S-ICD patients. These findings emphasize the importance of post-market registries in collaboration between clinicians and the manufacturer to optimize safety and efficacy in S-ICD treatment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrophysiology, Heart Center at the University of Cologne, Kerpener Str. 62, 50937 Cologne, Germany.
- CHU de Bordeaux, service de Cardiologie-électrophysiologie et stimulation cardiaque, INSERM, U 1045, F-33000 Bordeaux, France.
- Department of Cardiology, University Heart Center, University Hospital Zurich, Zurich, Switzerland.
- Cardiology, VCU Health Pauley Heart Center, Virginia Commonwealth University, Richmond, USA.
- Cardiac Rhythm Management Research Department, University Hospital Southampton NHS Foundation Trust, Southampton, United Kingdom of Great Britain & Northern Ireland.
- Department of Cardiology, University Heart & Vascular Center Hamburg, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
ORCID-Profile: Jonas Wörmann, Marc Strik, Kara Stout, Mohamed Elrefai, Nina Becher, Beat Schaer, Antonius van Stipdonk, Neil T Srinivasan, Sylvain Ploux, Alexander Breitenstein, Jordana Kron, Paul R Roberts, Tobias Toennis, Dominik Linz, Rajdip Dulai, Julia Hermes-Laufer, Jay Koneru, Ömer Erküner, Sebastian Dittrich, Jan-Hendrik van den Bruck, Jan-Hendrik Schipper, Arian Sultan, Kerstin D Rosenberger, Daniel Steven, Jakob Lüker
Zitieren
APA
Wörmann, J., Strik, M., Jurisic, S., Stout, K., Elrefai, M., Becher, N., Schaer, B., van Stipdonk, A., Srinivasan, N. T., Ploux, S., Breitenstein, A., Kron, J., Roberts, P. R., Toennis, T., Linz, D., Dulai, R., Hermes-Laufer, J., Koneru, J., Erküner, Ö., …, & Lüker, J. (2024). Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients. Europace. https://doi.org/10.1093/europace/euae170
Vancouver
Wörmann J, Strik M, Jurisic S, Stout K, Elrefai M, Becher N, et al. Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients. Europace. 2024. doi:10.1093/europace/euae170
BibTeX
@article{Wrmann2024,
author = {Wörmann, Jonas and Strik, Marc and Jurisic, Stjepan and Stout, Kara and Elrefai, Mohamed and Becher, Nina and Schaer, Beat and van Stipdonk, Antonius and Srinivasan, Neil T and Ploux, Sylvain and Breitenstein, Alexander and Kron, Jordana and Roberts, Paul R and Toennis, Tobias and Linz, Dominik and Dulai, Rajdip and Hermes-Laufer, Julia and Koneru, Jay and Erküner, Ömer and Dittrich, Sebastian and van den Bruck, Jan-Hendrik and Schipper, Jan-Hendrik and Sultan, Arian and Rosenberger, Kerstin D and Steven, Daniel and Lüker, Jakob},
title = {Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients},
journal = {Europace},
year = {2024},
doi = {10.1093/europace/euae170}
}
EUVIMED