Procedural outcome and follow-up of stylet-driven leads compared with lumenless leads for left bundle branch area pacing

Aarthiga Sritharan iD, Nikola Kozhuharov iD, Nicolas Masson, Elise Bakelants iD, Valérian Valiton iD, Haran Burri iD

Europace · 2023

DOI: 10.1093/europace/euad295 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Europace, 2023. Verfasst von Aarthiga Sritharan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Pacing Unit, Department of Cardiology, University Hospital of Geneva, Rue Gabrielle-Perret-Gentil 4, CH-1211 Genève, Switzerland. Hauptthemen laut MEDLINE: Bundle of His, Cardiac Pacing, Artificial, Bundle-Branch Block, Pacemaker, Artificial. Die Arbeit wurde bisher 55-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: GeCOR.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Left bundle branch area pacing (LBBAP) is most often delivered using lumenless leads (LLLs), but may also be performed using stylet-driven leads (SDLs). There are limited reports on the comparison of these tools, mainly limited to reports describing initial operator experience or without detailed procedural data. Our aim was to perform an in-depth comparison of SDLs and LLLs for LBBAP at implantation and follow-up in a larger cohort of patients with experience that extends beyond that of the initial learning curve. Methods and results A total of 306 consecutive patients (age 77 ± 11 years, 183 males) undergoing LBBAP implantation at a single centre were prospectively included. The population was split into two groups of 153 patients based on the initial use of an SDL (from 4 manufacturers) or an LLL. After having discounted the initial learning curve of 50 patients, there was no difference in the success rate between the initial use of lead type (96.0% with SDL vs. 94.3% with LLL, P = 0.56). There were no significant differences in success between lead models. Electrocardiogram and electrical parameters were comparable between the groups. Post-operative macro-dislodgement occurred in 4.3% of patients (essentially within the first day following implantation) and presumed micro-dislodgement with loss of conduction system capture or rise in threshold (occurring mostly during the first month) was observed in 4.7% of patients, without differences between groups.

Conclusion

Left bundle branch area pacing may be safely and effectively performed using either LLLs or SDLs, which provides implanters with alternatives for delivering this therapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bundle of HisCardiac Pacing, ArtificialBundle-Branch BlockPacemaker, ArtificialHumansMaleAgedFemaleTreatment OutcomeAged, 80 and overProspective StudiesEquipment DesignLearning CurveTime FactorsAction PotentialsFollow-Up Studies

Stichwörter der Autor:innen

ComplicationsConduction system pacingFailureLeft bundle branch area pacingLumenless leadStylet-driven leadSuccess

Autor:innen und Institutionen

  • Cardiac Pacing Unit, Department of Cardiology, University Hospital of Geneva, Rue Gabrielle-Perret-Gentil 4, CH-1211 Genève, Switzerland.

ORCID-Profile: Aarthiga Sritharan, Nikola Kozhuharov, Elise Bakelants, Valérian Valiton, Haran Burri

Förderung

  • GeCOR

Zitieren

APA

Sritharan, A., Kozhuharov, N., Masson, N., Bakelants, E., Valiton, V., & Burri, H. (2023). Procedural outcome and follow-up of stylet-driven leads compared with lumenless leads for left bundle branch area pacing. Europace. https://doi.org/10.1093/europace/euad295

Vancouver

Sritharan A, Kozhuharov N, Masson N, Bakelants E, Valiton V, Burri H. Procedural outcome and follow-up of stylet-driven leads compared with lumenless leads for left bundle branch area pacing. Europace. 2023. doi:10.1093/europace/euad295

BibTeX
@article{Sritharan2023,
  author = {Sritharan, Aarthiga and Kozhuharov, Nikola and Masson, Nicolas and Bakelants, Elise and Valiton, Valérian and Burri, Haran},
  title = {Procedural outcome and follow-up of stylet-driven leads compared with lumenless leads for left bundle branch area pacing},
  journal = {Europace},
  year = {2023},
  doi = {10.1093/europace/euad295}
}
Kostenlos starten →