Feasibility of three-dimensional artificial intelligence algorithm integration with intracardiac echocardiography for left atrial imaging during atrial fibrillation catheter ablation
Europace · 2023
DOI: 10.1093/europace/euad211 ↗Kurz gefasst
Fachartikel in der Zeitschrift Europace, 2023. Verfasst von Luigi Di Biase und 16 weiteren Autor:innen. Beteiligte Einrichtung: Montefiore-Einstein Center for Heart & Vascular Care, Department of Cardiology, Montefiore Medical Center, Albert Einstein College of Medicine, 111 E…. Hauptthemen laut MEDLINE: Atrial Fibrillation, Pulmonary Veins, Catheter Ablation. Die Arbeit wurde bisher 31-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Aims
Intracardiac echocardiography (ICE) is a useful but operator-dependent tool for left atrial (LA) anatomical rendering during atrial fibrillation (AF) ablation. The CARTOSOUND FAM Module, a new deep learning (DL) imaging algorithm, has the potential to overcome this limitation. This study aims to evaluate feasibility of the algorithm compared to cardiac computed tomography (CT) in patients undergoing AF ablation. Methods and results In 28 patients undergoing AF ablation, baseline patient information was recorded, and three-dimensional (3D) shells of LA body and anatomical structures [LA appendage/left superior pulmonary vein/left inferior pulmonary vein/right superior pulmonary vein/right inferior pulmonary vein (RIPV)] were reconstructed using the DL algorithm. The selected ultrasound frames were gated to end-expiration and max LA volume. Ostial diameters of these structures and carina-to-carina distance between left and right pulmonary veins were measured and compared with CT measurements. Anatomical accuracy of the DL algorithm was evaluated by three independent electrophysiologists using a three-anchor scale for LA anatomical structures and a five-anchor scale for LA body. Ablation-related characteristics were summarized. The algorithm generated 3D reconstruction of LA anatomies, and two-dimensional contours overlaid on ultrasound input frames. Average calculation time for LA reconstruction was 65 s. Mean ostial diameters and carina-to-carina distance were all comparable to CT without statistical significance. Ostial diameters and carina-to-carina distance also showed moderate to high correlation (r = 0.52–0.75) except for RIPV (r = 0.20). Qualitative ratings showed good agreement without between-rater differences. Average procedure time was 143.7 ± 43.7 min, with average radiofrequency time 31.6 ± 10.2 min. All patients achieved ablation success, and no immediate complications were observed.
Conclusion
DL algorithm integration with ICE demonstrated considerable accuracy compared to CT and qualitative physician assessment. The feasibility of ICE with this algorithm can potentially further streamline AF ablation workflow.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Montefiore-Einstein Center for Heart & Vascular Care, Department of Cardiology, Montefiore Medical Center, Albert Einstein College of Medicine, 111 E 210th street, Bronx, NY, USA.
- Biosense Webster, Inc., Irvine, CA, USA.
- Department of Medicine and Surgery, University of Insubria, Varese, Italy.
- St. David's Medical Center, Texas Cardiac Arrhythmia Institute, Austin, TX, USA.
ORCID-Profile: Luigi Di Biase, Fengwei Zou, Jacopo Marazzato, Domenico Della Rocca, Andrea Natale, Majd Al Deen Alhuarrat, Guy Haiman, David Haimovich, Richard A Matthew, Jaclyn Alcazar, Graça Costa, Xiaodong Zhang
Zitieren
APA
Di Biase, L., Zou, F., Lin, A. N., Grupposo, V., Marazzato, J., Tarantino, N., Della Rocca, D., Mohanty, S., Natale, A., Alhuarrat, M. A. D., Haiman, G., Haimovich, D., Matthew, R. A., Alcazar, J., Costa, G., Urman, R., & Zhang, X. (2023). Feasibility of three-dimensional artificial intelligence algorithm integration with intracardiac echocardiography for left atrial imaging during atrial fibrillation catheter ablation. Europace. https://doi.org/10.1093/europace/euad211
Vancouver
Di Biase L, Zou F, Lin AN, Grupposo V, Marazzato J, Tarantino N, et al. Feasibility of three-dimensional artificial intelligence algorithm integration with intracardiac echocardiography for left atrial imaging during atrial fibrillation catheter ablation. Europace. 2023. doi:10.1093/europace/euad211
BibTeX
@article{DiBiase2023,
author = {Di Biase, Luigi and Zou, Fengwei and Lin, Aung N and Grupposo, Vito and Marazzato, Jacopo and Tarantino, Nicola and Della Rocca, Domenico and Mohanty, Sanghamitra and Natale, Andrea and Alhuarrat, Majd Al Deen and Haiman, Guy and Haimovich, David and Matthew, Richard A and Alcazar, Jaclyn and Costa, Graça and Urman, Roy and Zhang, Xiaodong},