Comparison of novel ventricular pacing strategies using an electro-mechanical simulation platform
Europace · 2023
DOI: 10.1093/europace/euad144 ↗Kurz gefasst
Fachartikel in der Zeitschrift Europace, 2023. Verfasst von Roel Meiburg und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Universiteitssingel 40, 6200 MD, M…. Hauptthemen laut MEDLINE: Heart Ventricles, Ventricular Septum. Die Arbeit wurde bisher 49-mal zitiert (Crossref) und nennt 66 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Dutch Heart Foundation, Netherlands Organisation for Scientific Research, European Union's Horizon 2020 research and innovation programme.
Abstract (Originalsprache)
Aims
Focus of pacemaker therapy is shifting from right ventricular (RV) apex pacing (RVAP) and biventricular pacing (BiVP) to conduction system pacing. Direct comparison between the different pacing modalities and their consequences to cardiac pump function is difficult, due to the practical implications and confounding variables. Computational modelling and simulation provide the opportunity to compare electrical, mechanical, and haemodynamic consequences in the same virtual heart. Methods and results Using the same single cardiac geometry, electrical activation maps following the different pacing strategies were calculated using an Eikonal model on a three-dimensional geometry, which were then used as input for a lumped mechanical and haemodynamic model (CircAdapt). We then compared simulated strain, regional myocardial work, and haemodynamic function for each pacing strategy. Selective His-bundle pacing (HBP) best replicated physiological electrical activation and led to the most homogeneous mechanical behaviour. Selective left bundle branch (LBB) pacing led to good left ventricular (LV) function but significantly increased RV load. RV activation times were reduced in non-selective LBB pacing (nsLBBP), reducing RV load but increasing heterogeneity in LV contraction. LV septal pacing led to a slower LV and more heterogeneous LV activation than nsLBBP, while RV activation was similar. BiVP led to a synchronous LV–RV, but resulted in a heterogeneous contraction. RVAP led to the slowest and most heterogeneous contraction. Haemodynamic differences were small compared to differences in local wall behaviour.
Conclusion
Using a computational modelling framework, we investigated the mechanical and haemodynamic outcome of the prevailing pacing strategies in hearts with normal electrical and mechanical function. For this class of patients, nsLBBP was the best compromise between LV and RV function if HBP is not possible.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Universiteitssingel 40, 6200 MD, Maastricht, The Netherlands.
- Department of Cardiology, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center+ (MUMC+), Maastricht, The Netherlands.
- Department of Cardiovascular Diseases, Faculty of Medicine, Suez Canal University, Ismailia, Egypt.
- Department of Cardiovascular Sciences, Policlinico Casilino of Rome, Rome, Italy.
- Department of Cardiology, Radboud University Medical Centre (Radboudumc), Nijmegen, The Netherlands.
- Department of Physiology, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Maastricht, The Netherlands.
ORCID-Profile: Roel Meiburg, Jesse H J Rijks, Ahmed S Beela, Edoardo Bressi, Domenico Grieco, Tammo Delhaas, Frits W Prinzen, Kevin Vernooy, Joost Lumens
Förderung
- Dutch Heart Foundation
- Netherlands Organisation for Scientific Research
- European Union's Horizon 2020 research and innovation programme
- Marie Sklodowska-Curie
- European Union’s Horizon 2020 research and innovation programme
Zitieren
APA
Meiburg, R., Rijks, J. H. J., Beela, A. S., Bressi, E., Grieco, D., Delhaas, T., Luermans, J. G. L., Prinzen, F. W., Vernooy, K., & Lumens, J. (2023). Comparison of novel ventricular pacing strategies using an electro-mechanical simulation platform. Europace. https://doi.org/10.1093/europace/euad144
Vancouver
Meiburg R, Rijks JHJ, Beela AS, Bressi E, Grieco D, Delhaas T, et al. Comparison of novel ventricular pacing strategies using an electro-mechanical simulation platform. Europace. 2023. doi:10.1093/europace/euad144
BibTeX
@article{Meiburg2023,
author = {Meiburg, Roel and Rijks, Jesse H J and Beela, Ahmed S and Bressi, Edoardo and Grieco, Domenico and Delhaas, Tammo and Luermans, Justin G Lm and Prinzen, Frits W and Vernooy, Kevin and Lumens, Joost},
title = {Comparison of novel ventricular pacing strategies using an electro-mechanical simulation platform},
journal = {Europace},
year = {2023},
doi = {10.1093/europace/euad144}
}
EUVIMED