High-power chargers for electric vehicles: are they safe for patients with pacemakers and defibrillators?
Europace · 19.05.2023 · 25 (5)
DOI: 10.1093/europace/euad042 ↗Kurz gefasst
Fachartikel in der Zeitschrift Europace, online erschienen am 17.04.2023. Verfasst von Carsten Lennerz und 18 weiteren Autor:innen. Beteiligte Einrichtung: German Heart Centre Munich, Department of Electrophysiology, Technical University of Munich, Lazarettstr. 36, 80636 Munich, Germany. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Pacemaker, Artificial. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: German Foundation of Heart Research, German Social Accident Insurance, Deutsche Gesetzliche Unfallversicherung.
Abstract (Originalsprache)
Aims
Battery electric vehicle (BEV) sales and use are rapidly expanding. Battery electric vehicles, along with their charging stations, are a potential source of electromagnetic interference (EMI) for patients with cardiac implantable electronic devices (CIEDs). The new ‘high-power’ charging stations have the potential to create strong electromagnetic fields and induce EMI in CIEDs, and their safety has not been evaluated. Methods and results A total of 130 CIED patients performed 561 charges of four BEVs and a test vehicle (350 kW charge capacity) using high-power charging stations under continuous 6-lead electrocardiogram monitoring. The charging cable was placed directly over the CIED, and devices were programmed to maximize the chance of EMI detection. Cardiac implantable electronic devices were re-interrogated after patients charged all BEVs and the test vehicle for evidence of EMI. There were no incidences of EMI, specifically no over-sensing, pacing inhibition, inappropriate tachycardia detection, mode switching, or spontaneous reprogramming. The risk of EMI on a patient-based analysis is 0/130 [95% confidence interval (CI) 0%–2%], and the risk of EMI on a charge-based analysis is 0/561 (95% CI 0%–0.6%). The effective magnetic field along the charging cable was 38.65 µT and at the charging station was 77.9 µT.
Conclusions
The use of electric cars with high-power chargers by patients with cardiac devices appears to be safe with no evidence of clinically relevant EMI. Reasonable caution, by minimizing the time spent in close proximity with the charging cables, is still advised as the occurrence of very rare events cannot be excluded from our results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- German Heart Centre Munich, Department of Electrophysiology, Technical University of Munich, Lazarettstr. 36, 80636 Munich, Germany.
- DZHK (German Centre for Cardiovascular Research) partner site Munich Heart Alliance, Pettenkoferstr. 8a & 9, 80336 Munich, Germany.
- German Heart Centre Munich, Technical University of Munich, Lazarettstr. 36, 80636 Munich, Germany.
- Cardiology Department, Auckland City Hospital, 2 Park Road, Grafton, 1023 Auckland, New Zealand.
ORCID-Profile: Carsten Lennerz, Claudia Schaarschmidt, Patrick Blažek, Katharina Knoll, Marc Kottmaier, Tilko Reents, Miruna Popa, Katharina Wimbauer, Fabian Bahlke, Hannah Krafft, Florian Englert, Lena Friedrich, Heribert Schunkert, Isabel Deisenhofer, Matthew O'Connor
Förderung
- German Foundation of Heart Research
- German Social Accident Insurance
- Deutsche Gesetzliche Unfallversicherung
Zitieren
APA
Lennerz, C., Schaarschmidt, C., Blažek, P., Knoll, K., Kottmaier, M., Reents, T., Bourier, F., Lengauer, S., Popa, M., Wimbauer, K., Bahlke, F., Krafft, H., Englert, F., Friedrich, L., Schunkert, H., Hessling, G., Deisenhofer, I., Kolb, C., & O'Connor, M. (2023). High-power chargers for electric vehicles: are they safe for patients with pacemakers and defibrillators?. Europace, 25(5). https://doi.org/10.1093/europace/euad042
Vancouver
Lennerz C, Schaarschmidt C, Blažek P, Knoll K, Kottmaier M, Reents T, et al. High-power chargers for electric vehicles: are they safe for patients with pacemakers and defibrillators?. Europace. 2023;25(5). doi:10.1093/europace/euad042
BibTeX
@article{Lennerz2023,
author = {Lennerz, Carsten and Schaarschmidt, Claudia and Blažek, Patrick and Knoll, Katharina and Kottmaier, Marc and Reents, Tilko and Bourier, Felix and Lengauer, Sarah and Popa, Miruna and Wimbauer, Katharina and Bahlke, Fabian and Krafft, Hannah and Englert, Florian and Friedrich, Lena and Schunkert, Heribert and Hessling, Gabriele and Deisenhofer, Isabel and Kolb, Christof and O'Connor, Matthew},
title = {High-power chargers for electric vehicles: are they safe for patients with pacemakers and defibrillators?},
journal = {Europace},
year = {2023},
volume = {25},
number = {5},
doi = {10.1093/europace/euad042}
}
EUVIMED