Usefulness of delivery catheter on accurate right ventricular septal pacing: Mt FUJI trial
EP Europace · 2023
DOI: 10.1093/europace/euad027 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift EP Europace, 2023. Verfasst von Yoshihisa Naruse und 21 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, 1-20-1 Handayama, Higashi-ward, Hamamatsu 431-3192, Japan. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Ventricular Septum. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Medtronic Japan Co., Ltd.
Abstract (Originalsprache)
Aims
Although the delivery catheter system for pacemaker-lead implantation is a new alternative to the stylet system, no randomized controlled trial has addressed the difference in right ventricular (RV) lead placement accuracy to the septum between the stylet and the delivery catheter systems. This multicentre prospective randomized controlled trial aimed to prove the efficacy of the delivery catheter system for accurate delivery of RV lead to the septum. Methods and results In this trial, 70 patients (mean age 78 ± 11 years; 30 men) with pacemaker indications of atrioventricular block were randomized to the delivery catheter or the stylet groups. Right ventricular lead tip positions were assessed using cardiac computed tomography within 4 weeks of pacemaker implantation. Lead tip positions were classified into RV septum, anterior/posterior edge of the RV septal wall, and RV free wall. The primary endpoint was the success rate of RV lead tip placement to the RV septum.
Results
Right ventricular leads were implanted as per allocation in all patients. The delivery catheter group had higher success rate of RV lead deployment to the septum (78 vs. 50%; P = 0.024) and narrower paced QRS width (130 ± 19 vs. 142 ± 15 ms P = 0.004) than those in the stylet group. However, there was no significant difference in procedure time [91 (IQR 68–119) vs. 85 (59–118) min; P = 0.488] or the incidence of RV lead dislodgment (0 vs. 3%; P = 0.486).
Conclusion
The delivery catheter system can achieve a higher success rate of RV lead placement to the RV septum and narrower paced QRS width than the stylet system. Trial registration number jRCTs042200014 (https://jrct.niph.go.jp/en-latest-detail/jRCTs042200014)
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, 1-20-1 Handayama, Higashi-ward, Hamamatsu 431-3192, Japan.
- Department of Cardiology, Chutoen General Medical Center, 1-1 Shobugaike, Kakegawa, 436-0040, Japan.
- Department of Cardiology, Juntendo University Shizuoka Hospital, 1129 Nagaoka, Izunokuni, 410-2211, Japan.
- Department of Cardiovascular Medicine, Shizuoka Medical Center, 762-1 Nagasawa, Sunto-gun, 411-8611, Japan.
- Department of Cardiology, Seirei Mikatahara General Hospital, 3453 Mikatahara-cho, Kita-ward, Hamamatsu, 433-8558, Japan.
- Department of Cardiology, Seirei Hamamatsu General Hospital, 2-12-1 Sumiyoshi, Naka-ward, Hamamatsu, 430-0906, Japan.
ORCID-Profile: Yoshihisa Naruse, Makoto Sano, Shuji Morikawa, Tomoyuki Shiozawa, Shintaro Takano, Michio Ogano, Kei Kimura, Keisuke Miyajima, Ryuta Henmi, Masahiro Muto, Natsuko Hosoya, Hideyuki Hasebe, Akira Mizukami, Keisuke Iguchi, Akiko Atsumi
Förderung
- Medtronic Japan Co., Ltd
Zitieren
APA
Naruse, Y., Sano, M., Kurebayashi, N., Morikawa, S., Tsurumi, N., Shiozawa, T., Takano, S., Ogano, M., Kimura, K., Miyajima, K., Sugiura, R., Henmi, R., Muto, M., Hosoya, N., Hasebe, H., Mizukami, A., Iguchi, K., Atsumi, A., Odagiri, K., …, & Mt FUJI trial investigators (2023). Usefulness of delivery catheter on accurate right ventricular septal pacing: Mt FUJI trial. EP Europace. https://doi.org/10.1093/europace/euad027
Vancouver
Naruse Y, Sano M, Kurebayashi N, Morikawa S, Tsurumi N, Shiozawa T, et al. Usefulness of delivery catheter on accurate right ventricular septal pacing: Mt FUJI trial. EP Europace. 2023. doi:10.1093/europace/euad027
BibTeX
@article{Naruse2023,
author = {Naruse, Yoshihisa and Sano, Makoto and Kurebayashi, Nobutake and Morikawa, Shuji and Tsurumi, Naoki and Shiozawa, Tomoyuki and Takano, Shintaro and Ogano, Michio and Kimura, Kei and Miyajima, Keisuke and Sugiura, Ryo and Henmi, Ryuta and Muto, Masahiro and Hosoya, Natsuko and Hasebe, Hideyuki and Mizukami, Akira and Iguchi, Keisuke and Atsumi, Akiko and Odagiri, Keiichi and Kiyama, Yumi and Maekawa, Yuichiro and Mt FUJI trial investigators, Mt FUJI trial investigators},
title = {Usefulness of delivery catheter on accurate right ventricular septal pacing: Mt FUJI trial},
journal = {EP Europace},
year = {2023},
doi = {10.1093/europace/euad027}
}
EUVIMED