Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation
EP Europace · 2023
DOI: 10.1093/europace/euad011 ↗Kurz gefasst
Fachartikel in der Zeitschrift EP Europace, 2023. Verfasst von Astrid N L Hermans und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht, P. Debyelaan 25, 6229 HX Maastricht,…. Hauptthemen laut MEDLINE: Atrial Fibrillation. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aims
Although mobile health tools using photoplethysmography (PPG) technology have been validated for the detection of atrial fibrillation (AF), their utility for heart rate assessment during AF remains unclear. Therefore, we aimed to evaluate the accuracy of continuous PPG-based 1 min mean heart rate assessment during AF. Methods and results Persistent AF patients were provided with Holter electrocardiography (ECG) (for ≥24 h) simultaneously with a PPG-equipped smartwatch. Both the PPG-based smartwatch and Holter ECG automatically and continuously monitored patients’ heart rate/rhythm. ECG and PPG recordings were synchronized and divided into 1 min segments, from which a PPG-based and an ECG-based average heart rate estimation were extracted. In total, 47 661 simultaneous ECG and PPG 1 min heart rate segments were analysed in 50 patients (34% women, age 73 ± 8 years). The agreement between ECG-determined and PPG-determined 1 min mean heart rate was high [root mean squared error (RMSE): 4.7 bpm]. The 1 min mean heart rate estimated using PPG was accurate within ±10% in 93.7% of the corresponding ECG-derived 1 min mean heart rate segments. PPG-based 1 min mean heart rate estimation was more often accurate during night-time (97%) than day-time (91%, P < 0.001) and during low levels (96%) compared to high levels of motion (92%, P < 0.001). A neural network with a 10 min history of the recording did not further improve the PPG-based 1 min mean heart rate assessment [RMSE: 4.4 (95% confidence interval: 3.5–5.2 bpm)]. Only chronic heart failure was associated with a lower agreement between ECG-derived and PPG-derived 1 min mean heart rates (P = 0.040).
Conclusion
During persistent AF, continuous PPG-based 1 min mean heart rate assessment is feasible in 60% of the analysed period and shows high accuracy compared with Holter ECG for heart rates <110 bpm.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht, P. Debyelaan 25, 6229 HX Maastricht, The Netherlands.
- Department of Biomedical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Nørregade 10, 1165 Copenhagen, Denmark.
- Institute of Pharmacology, West German Heart and Vascular Centre, University Duisburg-Essen, Forsthausweg 2, 47057 Duisburg, Germany.
- 1st Department of Cardiology, Medical University of Warsaw, Żwirki i Wigury 61, 02-091 Warsaw, Poland.
- Qompium NV, Kempische Steenweg 293/16, 3500 Hasselt, Belgium.
- Department of Cardiology, Radboud University Medical Centre, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
ORCID-Profile: Astrid N L Hermans, Jonas L Isaksen, Monika Gawalko, Nikki A H A Pluymaekers, Rachel M J van der Velden, Justin G L M Luermans, Martin Manninger, Joost Lumens, Jørgen K Kanters, Dominik Linz
Zitieren
APA
Hermans, A. N. L., Isaksen, J. L., Gawalko, M., Pluymaekers, N. A. H. A., van der Velden, R. M. J., Snippe, H., Evens, S., De Witte, G., Luermans, J. G. L. M., Manninger, M., Lumens, J., Kanters, J. K., & Linz, D. (2023). Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation. EP Europace. https://doi.org/10.1093/europace/euad011
Vancouver
Hermans ANL, Isaksen JL, Gawalko M, Pluymaekers NAHA, van der Velden RMJ, Snippe H, et al. Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation. EP Europace. 2023. doi:10.1093/europace/euad011
BibTeX
@article{Hermans2023,
author = {Hermans, Astrid N L and Isaksen, Jonas L and Gawalko, Monika and Pluymaekers, Nikki A H A and van der Velden, Rachel M J and Snippe, Hilco and Evens, Stijn and De Witte, Glenn and Luermans, Justin G L M and Manninger, Martin and Lumens, Joost and Kanters, Jørgen K and Linz, Dominik},
title = {Accuracy of continuous photoplethysmography-based 1 min mean heart rate assessment during atrial fibrillation},
journal = {EP Europace},
year = {2023},
doi = {10.1093/europace/euad011}
}
EUVIMED