Comparison of methods for delivering cardiac resynchronization therapy: an acute electrical and haemodynamic within-patient comparison of left bundle branch area, His bundle, and biventricular pacing
EP Europace · 2023
DOI: 10.1093/europace/euac245 ↗Kurz gefasst
Kommentar in der Zeitschrift EP Europace, 2023. Verfasst von Nadine Ali und 18 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, National Heart and Lung Institute, Imperial College London, Hammersmith Hospital Du Cane Road, London W120HS, UK. Hauptthemen laut MEDLINE: Cardiac Resynchronization Therapy, Heart Failure. Die Arbeit wurde bisher 61-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: British Heart Foundation.
Abstract (Originalsprache)
Aims
Left bundle branch area pacing (LBBAP) is a promising method for delivering cardiac resynchronization therapy (CRT), but its relative physiological effectiveness compared with His bundle pacing (HBP) is unknown. We conducted a within-patient comparison of HBP, LBBAP, and biventricular pacing (BVP). Methods and results Patients referred for CRT were recruited. We assessed electrical response using non-invasive mapping, and acute haemodynamic response using a high-precision haemodynamic protocol. Nineteen patients were recruited: 14 male, mean LVEF of 30%. Twelve had time for BVP measurements. All three modalities reduced total ventricular activation time (TVAT), (ΔTVATHBP -43 ± 14 ms and ΔTVATLBBAP −35 ± 20 ms vs. ΔTVATBVP −19 ± 30 ms, P = 0.03 and P = 0.1, respectively). HBP produced a significantly greater reduction in TVAT compared with LBBAP in all 19 patients (−46 ± 15 ms, −36 ± 17 ms, P = 0.03). His bundle pacing and LBBAP reduced left ventricular activation time (LVAT) more than BVP (ΔLVATHBP −43 ± 16 ms, P < 0.01 vs. BVP, ΔLVATLBBAP −45 ± 17 ms, P < 0.01 vs. BVP, ΔLVATBVP −13 ± 36 ms), with no difference between HBP and LBBAP (P = 0.65). Acute systolic blood pressure was increased by all three modalities. In the 12 with BVP, greater improvement was seen with HBP and LBBAP (6.4 ± 3.8 mmHg BVP, 8.1 ± 3.8 mmHg HBP, P = 0.02 vs. BVP and 8.4 ± 8.2 mmHg for LBBAP, P = 0.3 vs. BVP), with no difference between HBP and LBBAP (P = 0.8).
Conclusion
HBP delivered better ventricular resynchronization than LBBAP because right ventricular activation was slower during LBBAP. But LBBAP was not inferior to HBP with respect to LV electrical resynchronization and acute haemodynamic response.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, National Heart and Lung Institute, Imperial College London, Hammersmith Hospital Du Cane Road, London W120HS, UK.
- Medtronic Limited, Building 9, Croxley Green Business Park, Watford WD18 8WW, UK.
- Department of Cardiology, St Richards Hospital, University Hospitals Sussex NHS Foundation Trust., Spitalfield Ln, Chichester PO19 6SE, UK.
- Department of Cardiology, St Bartholomew's Hospital and North Middlesex University Hospital, W Smithfield, London EC1A 7BE, UK.
ORCID-Profile: Nadine Ali, Ahran D Arnold, Alejandra A Miyazawa, Daniel Keene, Ji-Jian Chow, Nicholas S Peters, Prapa Kanagaratnam, Norman Qureshi, Fu Siong Ng, Nick W F Linton, David C Lefroy, Darrel P Francis, Lim Phang Boon, Mark A Tanner, Matthew J Shun-Shin, Graham D Cole, Zachary I Whinnett
Förderung
- British Heart Foundation
Zitieren
APA
Ali, N., Arnold, A. D., Miyazawa, A. A., Keene, D., Chow, J. J., Little, I., Peters, N. S., Kanagaratnam, P., Qureshi, N., Ng, F. S., Linton, N. W. F., Lefroy, D. C., Francis, D. P., Phang Boon, L., Tanner, M. A., Muthumala, A., Shun-Shin, M. J., Cole, G. D., & Whinnett, Z. I. (2023). Comparison of methods for delivering cardiac resynchronization therapy: an acute electrical and haemodynamic within-patient comparison of left bundle branch area, His bundle, and biventricular pacing. EP Europace. https://doi.org/10.1093/europace/euac245
Vancouver
Ali N, Arnold AD, Miyazawa AA, Keene D, Chow JJ, Little I, et al. Comparison of methods for delivering cardiac resynchronization therapy: an acute electrical and haemodynamic within-patient comparison of left bundle branch area, His bundle, and biventricular pacing. EP Europace. 2023. doi:10.1093/europace/euac245
BibTeX
@article{Ali2023,
author = {Ali, Nadine and Arnold, Ahran D and Miyazawa, Alejandra A and Keene, Daniel and Chow, Ji-Jian and Little, Ian and Peters, Nicholas S and Kanagaratnam, Prapa and Qureshi, Norman and Ng, Fu Siong and Linton, Nick W F and Lefroy, David C and Francis, Darrel P and Phang Boon, Lim and Tanner, Mark A and Muthumala, Amal and Shun-Shin, Matthew J and Cole, Graham D and Whinnett, Zachary I},
title = {Comparison of methods for delivering cardiac resynchronization therapy: an acute electrical and haemodynamic within-patient comparison of left bundle branch area, His bundle, and biventricular pacing},
journal = {EP Europace},
year = {2023},
doi = {10.1093/europace/euac245}
}
EUVIMED