A head-to-head comparison of laser vs. powered mechanical sheaths as first choice and second line extraction tools
EP Europace · 2023
DOI: 10.1093/europace/euac200 ↗Kurz gefasst
Fachartikel in der Zeitschrift EP Europace, 2023. Verfasst von Elod-Janos Zsigmond und 8 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Electrophysiology Division, Department of Internal Medicine, University of Szeged, Semmelweis str. 8, 6725 Szeged, Hungary. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Pacemaker, Artificial. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Aims
During transvenous lead extraction (TLE) longer dwelling time often requires the use of powered sheaths. This study aimed to compare outcomes with the laser and powered mechanical tools. Methods and results Single-centre data from consecutive patients undergoing TLE between 2012 and 2021 were retrospectively analysed. Efficacy and safety of the primary extraction tool were compared. Procedures requiring crossover between powered sheaths were also analysed. Moreover, we examined the efficacy of each level of the stepwise approach. Out of 166 patients, 142 (age 65.4 ± 13.7 years) underwent TLE requiring advanced techniques with 245 leads (dwelling time 9.4 ± 6.3 years). Laser sheaths were used in 64.9%, powered mechanical sheaths in 35.1% of the procedures as primary extraction tools. Procedural success rate was 85.5% with laser and 82.5% with mechanical sheaths (P = 0.552). Minor and major complications were observed in similar rate. Procedural mortality occurred only in the laser group in the case of three patients. Crossover was needed in 19.5% after laser and in 12.8% after mechanical extractions (P = 0.187). Among crossover procedures, only clinical success favoured the secondary mechanical arm (87.1 vs. 54.5%, aOR: 0.09, 95% CI: 0.01–0.79, P = 0.030). After step-by-step efficacy analysis, procedural success was 64.9% with the first-line extraction tool, 75.1% after crossover, 84.5% with bailout femoral snare, and 91.8% by non-emergency surgery.
Conclusion
The efficacy and safety of laser and mechanical sheaths were similar, however in the subgroup of crossover procedures mechanical tools had better performance regarding clinical success. Device diversity seems to help improving outcomes, especially in the most complicated cases.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiac Electrophysiology Division, Department of Internal Medicine, University of Szeged, Semmelweis str. 8, 6725 Szeged, Hungary.
- Heart Surgery Department, University of Szeged, Semmelweis str. 8, 6725 Szeged, Hungary.
- Department of Anaesthesiology and Intensive Therapy, University of Szeged, Semmelweis str. 6, 6725 Szeged, Hungary.
ORCID-Profile: Elod-Janos Zsigmond, Attila Benak, Endre Alacs, Szilvia Agocs, Mate Vamos
Zitieren
APA
Zsigmond, E. J., Saghy, L., Benak, A., Miklos, M., Makai, A., Hegedus, Z., Alacs, E., Agocs, S., & Vamos, M. (2023). A head-to-head comparison of laser vs. powered mechanical sheaths as first choice and second line extraction tools. EP Europace. https://doi.org/10.1093/europace/euac200
Vancouver
Zsigmond EJ, Saghy L, Benak A, Miklos M, Makai A, Hegedus Z, et al. A head-to-head comparison of laser vs. powered mechanical sheaths as first choice and second line extraction tools. EP Europace. 2023. doi:10.1093/europace/euac200
BibTeX
@article{Zsigmond2023,
author = {Zsigmond, Elod-Janos and Saghy, Laszlo and Benak, Attila and Miklos, Marton and Makai, Attila and Hegedus, Zoltan and Alacs, Endre and Agocs, Szilvia and Vamos, Mate},
title = {A head-to-head comparison of laser vs. powered mechanical sheaths as first choice and second line extraction tools},
journal = {EP Europace},
year = {2023},
doi = {10.1093/europace/euac200}
}
EUVIMED