Predictors of stroke volume improvement with AV-optimized conduction system pacing in patients with AV dromotropathy

Anja Zupan Mežnar, Tadej Žlahtič, Miha Mrak, Maja Ivanovski, David Žižek iD

ESC Heart Failure · 2026

DOI: 10.1093/eschf/xvag060 ↗

Kurz gefasst

Fachartikel in der Zeitschrift ESC Heart Failure, 2026. Verfasst von Anja Zupan Mežnar und 4 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine, University of Ljubljana, Vrazov trg 2, 1000 Ljubljana, Slovenia. Hauptthemen laut MEDLINE: Stroke Volume, Atrioventricular Block, Cardiac Pacing, Artificial, Heart Conduction System. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University Medical Centre Ljubljana.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Patients with first-degree atrioventricular (AV) block and mechanical AV dyssynchrony can present with heart failure (HF)-like symptoms. AV-optimized conduction system pacing (CSP) can improve haemodynamics and symptoms, but selection criteria remain uncertain. We aimed to identify electrocardiographic and echocardiographic predictors of an acute haemodynamic response to AV-optimized CSP in symptomatic first-degree AV block. Methods and Results Nineteen patients (mean age 60.5 ± 21.1 years; 37% female) with symptomatic first-degree AV block underwent baseline electrocardiography and echocardiography followed by AV-optimized conduction system pacing and repeat echocardiographic assessment. Electrocardiographic parameters (PR interval, P wave duration/PR interval ratio) and echocardiographic indices (E/A wave confluence, A-Q interval, and DFT/RR ratio) were tested for association with change in left ventricular stroke volume (LVSV). The mean PR interval was 395 ± 61 ms, the mean A-Q interval 155 ± 65 ms, and the mean DFT/RR ratio 0.34 ± 0.1. E/A wave confluence was present in 15 patients (79%). AV-optimized pacing increased LVSV by 7.8 ± 3.9 ml, corresponding to an 11.8 ± 5.7% relative increase (P < .01). Echocardiographic parameters were associated with LVSV response, including A-Q interval (r = 0.63, P = .004), DFT/RR ratio (r = −0.59, P = .008), and E/A wave confluence (r = 0.57, P = .01). Electrocardiographic parameters were not associated with LVSV change.

Conclusions

Echocardiographically assessed mechanical AV dyssynchrony, rather than electrocardiographic parameters, is associated with an acute haemodynamic response to pacing. Echocardiographic evaluation may help identify patients with prolonged PR interval who could benefit from AV-optimized CSP.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Stroke VolumeAtrioventricular BlockCardiac Pacing, ArtificialHeart Conduction SystemVentricular Function, LeftHeart VentriclesHumansFemaleElectrocardiographyEchocardiographyMiddle AgedMaleAgedFollow-Up Studies

Stichwörter der Autor:innen

Atrioventricular conduction blockAtrioventricular dyssynchronyAtrioventricular optimizationConduction system pacing

Autor:innen und Institutionen

  • Faculty of Medicine, University of Ljubljana, Vrazov trg 2, 1000 Ljubljana, Slovenia.
  • Department of Cardiology, University Medical Centre Ljubljana, Zaloška 2, 1000 Ljubljana, Slovenia.

ORCID-Profile: David Žižek

Förderung

  • University Medical Centre Ljubljana

Zitieren

APA

Zupan Mežnar, A., Žlahtič, T., Mrak, M., Ivanovski, M., & Žižek, D. (2026). Predictors of stroke volume improvement with AV-optimized conduction system pacing in patients with AV dromotropathy. ESC Heart Failure. https://doi.org/10.1093/eschf/xvag060

Vancouver

Zupan Mežnar A, Žlahtič T, Mrak M, Ivanovski M, Žižek D. Predictors of stroke volume improvement with AV-optimized conduction system pacing in patients with AV dromotropathy. ESC Heart Failure. 2026. doi:10.1093/eschf/xvag060

BibTeX
@article{ZupanMenar2026,
  author = {Zupan Mežnar, Anja and Žlahtič, Tadej and Mrak, Miha and Ivanovski, Maja and Žižek, David},
  title = {Predictors of stroke volume improvement with AV-optimized conduction system pacing in patients with AV dromotropathy},
  journal = {ESC Heart Failure},
  year = {2026},
  doi = {10.1093/eschf/xvag060}
}
Kostenlos starten →