A Novel Aortic End-to-End Anastomotic Stapler Device—Results of a Human Cadaver Study
European Journal of Cardio-Thoracic Surgery · 06.01.2026 · 68 (1)
DOI: 10.1093/ejcts/ezag015 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Cardio-Thoracic Surgery, online erschienen am 22.01.2026. Verfasst von Emilio Osorio-Jaramillo und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, A-1090 Vienna, Austria. Hauptthemen laut MEDLINE: Anastomosis, Surgical, Aorta, Blood Vessel Prosthesis Implantation, Surgical Staplers. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Aortic Technologies GmbH.
Abstract (Originalsprache)
Objectives
To evaluate a novel aortic anastomotic device designed for end-to-end anastomosis between a vascular prosthesis and the native aorta, aiming to facilitate aortic replacement and shorten circulatory arrest times.
Methods
Ascending aortic replacements with Dacron grafts and external Teflon felt strips were performed in 10 fresh human cadavers. Five procedures were performed using a continuous 4-0 polypropylene running suture, and 5 with the aortic anastomotic device. The device deploys single straight needle pins through the aortic wall and felt, securing them externally with a cap. Explanted graft-aorta specimens were connected to a pressure pump filled with glycerol/H2O solution (170/90 mm Hg, 5 minutes). The primary end-point was anastomotic time; secondary end-points were fluid loss and tissue trauma assessed macroscopically and microscopically.
Results
Median time for end-to-end anastomosis was significantly shorter with the aortic anastomotic device (05:39 [interquartile range, IQR 04:22-06:30] minutes:seconds) compared to conventional suture (09:17 [IQR 06:53-10:34]; P = .016). The number of pins per anastomosis was lower than the number of stitches (17 [15-18] vs 22 [21-24]; P = .008). Fluid loss was comparable between groups (100 [52.5-174.5] mL vs 45 [21.8-104.0] mL; P = .310). Microscopy revealed intimal tears in all sutured specimens but in none of the device group, P = .008.
Conclusions
In this preliminary human cadaver study, the novel aortic anastomotic device proved to be technically feasible, significantly faster than conventional suturing, and produced less tissue trauma while maintaining comparable sealing. The device may shorten circulatory arrest times and improve anastomotic quality in aortic surgery. Further in vivo studies are warranted. Clinical trial registration This was a prospective experimental cadaver study, not eligible for registration in a public registry for clinical trials.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, A-1090 Vienna, Austria.
- Division of Anatomy, Center for Anatomy and Cell Biology, Medical University of Vienna, A-1090 Vienna, Austria.
- National Cerebral and Cardiovascular Center, 564-8565 Suita, Osaka, Japan.
- Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, A-1090 Vienna, Austria.
ORCID-Profile: Emilio Osorio-Jaramillo, Paata Pruidze, Giorgi Didava, Wolfgang J Weninger, Daniel Zimpfer
Förderung
- Aortic Technologies GmbH
Zitieren
APA
Osorio-Jaramillo, E., El-Nashar, J., Kainuma, S., Unger, E., Schreiner, W., El Din, M., Pruidze, P., Didava, G., Weninger, W. J., Zimpfer, D., & Ehrlich, M. P. (2026). A Novel Aortic End-to-End Anastomotic Stapler Device—Results of a Human Cadaver Study. European Journal of Cardio-Thoracic Surgery, 68(1). https://doi.org/10.1093/ejcts/ezag015
Vancouver
Osorio-Jaramillo E, El-Nashar J, Kainuma S, Unger E, Schreiner W, El Din M, et al. A Novel Aortic End-to-End Anastomotic Stapler Device—Results of a Human Cadaver Study. European Journal of Cardio-Thoracic Surgery. 2026;68(1). doi:10.1093/ejcts/ezag015
BibTeX
@article{OsorioJaramillo2026,
author = {Osorio-Jaramillo, Emilio and El-Nashar, Jasmine and Kainuma, Satoshi and Unger, Ewald and Schreiner, Wilhelm and El Din, Mohamed and Pruidze, Paata and Didava, Giorgi and Weninger, Wolfgang J and Zimpfer, Daniel and Ehrlich, Marek P},
title = {A Novel Aortic End-to-End Anastomotic Stapler Device—Results of a Human Cadaver Study},
journal = {European Journal of Cardio-Thoracic Surgery},
year = {2026},
volume = {68},
number = {1},
doi = {10.1093/ejcts/ezag015}
}
EUVIMED