Automated Suture Securing Technology in Mitral Valve Surgery: A Strategy to Reduce Prosthetic Dehiscence?
European Journal of Cardio-Thoracic Surgery · 2026
DOI: 10.1093/ejcts/ezag013 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Cardio-Thoracic Surgery, 2026. Verfasst von Amila Kahrovic und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, 1090 Vienna, Austria. Hauptthemen laut MEDLINE: Mitral Valve, Suture Techniques, Heart Valve Prosthesis Implantation, Mitral Valve Annuloplasty. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
This study aimed to assess long-term outcomes of automated titanium fasteners versus hand-tied knots in mitral valve surgery.
Methods
In this retrospective, single-centre analysis, 2678 adult patients who underwent mitral valve repair or replacement between November 2008 and November 2024 at the Medical University of Vienna were included. Patients were grouped according to the suture-securing technique used: automated titanium fasteners versus hand-tied knots. The primary endpoint was prosthetic dehiscence (either mitral annuloplasty ring or valve replacement prosthesis) requiring reintervention. Secondary endpoints comprised ischaemic stroke, intracranial bleeding, and all-cause mortality during the follow-up period.
Results
Among the study population, 1072 (40%) underwent mitral valve surgery using an automated titanium fastener device, and 1606 (60%) with conventional hand-tied sutures. A total of 31 patients (1.2%) had prosthetic dehiscence during the follow-up period. The risk of prosthetic dehiscence was significantly lower in the automated titanium fastener group in both univariable (crude sub-hazard ratio [sHR] 0.32; 95% confidence interval [CI], 0.12-0.86, P = .023) and multivariable competing risk regression analysis (adjusted sHR 0.34; 95% CI, 0.12-0.91, P = .033). Automated titanium fastener group was not associated with an increased risk of ischaemic stroke (adjusted sHR 0.92; 95% CI, 0.67-1.27, P = .600), intracranial bleeding (adjusted sHR 0.89; 95% CI, 0.52-1.52, P = .675), or all-cause mortality (adjusted hazard ratio 0.93; 95% CI, 0.74-1.18, P = .559).
Conclusions
The use of an automated titanium fastener device seems to be associated with a lower risk of prosthetic dehiscence in mitral valve surgery. Due to the limited number of prosthetic dehiscence events and the potential for residual confounding, the results should be interpreted with caution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, 1090 Vienna, Austria.
- Department of Emergency Medicine, Medical University of Vienna, 1090 Vienna, Austria.
- Department of Infectious Diseases and Tropical Medicine, Medical University of Vienna, 1090 Vienna, Austria.
- Department of Cardiac Surgery, Medical University of Graz, A-8036 Graz, Austria.
ORCID-Profile: Amila Kahrovic, Paul Werner, Philipp Angleitner, Iuliana Coti, Kira Osipenko, Alfred Kocher, Daniel Zimpfer
Zitieren
APA
Kahrovic, A., Herkner, H., Werner, P., Angleitner, P., Coti, I., Osipenko, K., Lagler, H., Kocher, A., Ehrlich, M., Zimpfer, D., & Andreas, M. (2026). Automated Suture Securing Technology in Mitral Valve Surgery: A Strategy to Reduce Prosthetic Dehiscence?. European Journal of Cardio-Thoracic Surgery. https://doi.org/10.1093/ejcts/ezag013
Vancouver
Kahrovic A, Herkner H, Werner P, Angleitner P, Coti I, Osipenko K, et al. Automated Suture Securing Technology in Mitral Valve Surgery: A Strategy to Reduce Prosthetic Dehiscence?. European Journal of Cardio-Thoracic Surgery. 2026. doi:10.1093/ejcts/ezag013
BibTeX
@article{Kahrovic2026,
author = {Kahrovic, Amila and Herkner, Harald and Werner, Paul and Angleitner, Philipp and Coti, Iuliana and Osipenko, Kira and Lagler, Heimo and Kocher, Alfred and Ehrlich, Marek and Zimpfer, Daniel and Andreas, Martin},
title = {Automated Suture Securing Technology in Mitral Valve Surgery: A Strategy to Reduce Prosthetic Dehiscence?},
journal = {European Journal of Cardio-Thoracic Surgery},
year = {2026},
doi = {10.1093/ejcts/ezag013}
}
EUVIMED