EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Hydrodynamic ex vivo analysis of valve-sparing techniques: assessment and comparison

Sofia Di Leonardo iD, Danila Vella iD, Carmelo Savio Grillo, Carla Martorana, Salvatore Torre iD, Vincenzo Argano, Gaetano Burriesci iD

European Journal of Cardio-Thoracic Surgery · 2023

DOI: 10.1093/ejcts/ezad040 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Journal of Cardio-Thoracic Surgery, 2023. Verfasst von Sofia Di Leonardo und 6 weiteren Autor:innen. Beteiligte Einrichtung: Bioengineering Group, Ri.MED Foundation, Palermo, Italy. Hauptthemen laut MEDLINE: Sinus of Valsalva, Aortic Valve Insufficiency. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES Valve-sparing procedures are surgical techniques allowing to restore adequate function of the native aortic valve by replacing the dysfunctional ascending aorta with a prosthetic conduit. A number of techniques are currently used, such as Yacoub’s remodelling and David’s reimplantation, based on a regular straight conduit. More recently, the De Paulis proposed the use of bulging conduits to reconstruct the shape of the Valsalva sinuses. This work investigates the impact of the valve-sparing technique on the aortic valve function. METHODS The performance of 3 porcine aortic roots (Medtronic Freestyle™) was assessed in a cardiovascular pulse duplicator before and after performing 3 alternative valve-sparing procedures: David’s reimplantation, Yacoub’s remodelling and De Paulis’ reimplantation. RESULTS The porcine aortic roots, representative of the healthy native configuration, were characterized by the highest efficiency, with a mean energetic dissipation under normal operating conditions of 26 mJ. David’s and Yacoub’s techniques resulted in significantly lower performance (with mean energetic loss of about 70 mJ for both cases). The De Paulis’ procedure exhibited intermediate behaviour, with superior systolic performance and valve dynamics similar to the native case, and a mean energetic loss of 38 mJ. CONCLUSIONS The dynamics and performance after valve-sparing strongly depend on the adopted technique, with the use of conduits replicating the presence of Valsalva sinuses restoring more physiological conditions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Sinus of ValsalvaAortic Valve InsufficiencySwineAnimalsHydrodynamicsAortaAortic Valve

Stichwörter der Autor:innen

Ex vivo analysisAortic root prosthesisHydrodynamic performanceValsalva sinusesValve-sparing implants

Autor:innen und Institutionen

  • Bioengineering Group, Ri.MED Foundation, Palermo, Italy.
  • School of Medicine and Surgery, University of Palermo, Palermo, Italy.
  • Cardiac Surgery Unit, Policlinico Paolo Giaccone, University of Palermo, Palermo, Italy.
  • UCL Mechanical Engineering, University College London, London, UK.

ORCID-Profile: Sofia Di Leonardo, Danila Vella, Salvatore Torre, Gaetano Burriesci

Zitieren

APA

Di Leonardo, S., Vella, D., Grillo, C. S., Martorana, C., Torre, S., Argano, V., & Burriesci, G. (2023). Hydrodynamic ex vivo analysis of valve-sparing techniques: assessment and comparison. European Journal of Cardio-Thoracic Surgery. https://doi.org/10.1093/ejcts/ezad040

Vancouver

Di Leonardo S, Vella D, Grillo CS, Martorana C, Torre S, Argano V, et al. Hydrodynamic ex vivo analysis of valve-sparing techniques: assessment and comparison. European Journal of Cardio-Thoracic Surgery. 2023. doi:10.1093/ejcts/ezad040

BibTeX
@article{DiLeonardo2023,
  author = {Di Leonardo, Sofia and Vella, Danila and Grillo, Carmelo Savio and Martorana, Carla and Torre, Salvatore and Argano, Vincenzo and Burriesci, Gaetano},
  title = {Hydrodynamic ex vivo analysis of valve-sparing techniques: assessment and comparison},
  journal = {European Journal of Cardio-Thoracic Surgery},
  year = {2023},
  doi = {10.1093/ejcts/ezad040}
}
Kostenlos starten →