Technical limitations and error propagation in the 3D printing pipeline for congenital heart disease: an integrative review—a narrative synthesis

Ayush Balaji iD, Sinling Tiffany Yu, Nabil Hussein, Tristan Ramcharan iD, Benjamin Pinkey, Antonio Fratini, Adrian Crucean

European Heart Journal - Imaging Methods and Practice · 16.10.2026 · 4 (4)

DOI: 10.1093/ehjimp/qyag170 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Heart Journal - Imaging Methods and Practice, online erschienen am 05.11.2026. Verfasst von Ayush Balaji und 6 weiteren Autor:innen. Beteiligte Einrichtung: Birmingham Women's and Children's NHS Foundation Trust, Birmingham B4 6NH, UK. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Three-dimensional (3D) printing has gained prominence in the management of congenital heart disease (CHD) by augmenting spatial understanding, facilitating preoperative planning, and supporting education for clinicians and families. The clinical validity of printed models, however, is contingent upon accuracy maintained across a multistage pipeline that spans image acquisition, segmentation and modelling, printer-specific processing, and post-processing with subsequent clinical interpretation. Technical and human factors introduce discrepancies at each stage; these inaccuracies may accumulate and propagate, thereby altering the geometric and qualitative fidelity of the final model. This integrative review examines the principal sources of error across the pipeline, outlines mechanisms of propagation, and considers implications for decision-making in CHD, with particular emphasis on paediatric applications where permissible tolerances are narrow. Drawing upon contemporary literature, we provide recommendations for protocol standardization, validation procedures, operator training, and transparent documentation. An inter-observer experiment is proposed to quantify segmentation-induced variability using similarity and surface distance metrics, linking measurable divergence to potential clinical impact. A pragmatic error-minimizing pipeline is also described to improve reliability and promote reproducibility in clinical practice.

Schlagwörter

Stichwörter der Autor:innen

3D printingcongenital heart diseasegeometric accuracymedical image segmentationquality assurancesurgical planning

Autor:innen und Institutionen

  • Birmingham Women's and Children's NHS Foundation Trust, Birmingham B4 6NH, UK.
  • Hull York Medical School, University of Hull, Hull, UK.
  • Aston University, Birmingham, UK.

ORCID-Profile: Ayush Balaji, Tristan Ramcharan

Zitieren

APA

Balaji, A., Yu, S. T., Hussein, N., Ramcharan, T., Pinkey, B., Fratini, A., & Crucean, A. (2026). Technical limitations and error propagation in the 3D printing pipeline for congenital heart disease: an integrative review—a narrative synthesis. European Heart Journal - Imaging Methods and Practice, 4(4). https://doi.org/10.1093/ehjimp/qyag170

Vancouver

Balaji A, Yu ST, Hussein N, Ramcharan T, Pinkey B, Fratini A, et al. Technical limitations and error propagation in the 3D printing pipeline for congenital heart disease: an integrative review—a narrative synthesis. European Heart Journal - Imaging Methods and Practice. 2026;4(4). doi:10.1093/ehjimp/qyag170

BibTeX
@article{Balaji2026,
  author = {Balaji, Ayush and Yu, Sinling Tiffany and Hussein, Nabil and Ramcharan, Tristan and Pinkey, Benjamin and Fratini, Antonio and Crucean, Adrian},
  title = {Technical limitations and error propagation in the 3D printing pipeline for congenital heart disease: an integrative review—a narrative synthesis},
  journal = {European Heart Journal - Imaging Methods and Practice},
  year = {2026},
  volume = {4},
  number = {4},
  doi = {10.1093/ehjimp/qyag170}
}
Kostenlos starten →