Mixed coronary plaque phantom analysis by photon-counting CT: impact of calcium and iodine on low-attenuation plaque detection
European Heart Journal - Imaging Methods and Practice · 02.07.2026 · 4 (3)
DOI: 10.1093/ehjimp/qyag119 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Heart Journal - Imaging Methods and Practice, online erschienen am 22.07.2026. Verfasst von Bálint Szilveszter und 12 weiteren Autor:innen. Beteiligte Einrichtung: Heart and Vascular Center, Semmelweis University, Városmajor Street 68, Budapest 1122, Hungary. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union, János Bolyai Research Scholarship, STARTING GRANT.
Abstract (Originalsprache)
Aims
To test whether photon-counting detector CT (PCD-CT) improves low-attenuation plaque (LAP) detection compared with energy-integrating detector CT (EID-CT) using a mixed-plaque phantom and full-rotation helical images as high-resolution reference. Methods and results We imaged a custom phantom housing three different sized mixed-plaque inserts (LAP5.5mm, LAP4mm, and LAP7mm) combining iodinated lumen (600/1000 HU), calcified arcs (200/800 mg/cc), and a narrow LAP central strip (1–1.5 mm, 75 HU), embedded in a fat-equivalent ring (−60 HU). Polychromatic and virtual monoenergetic images (40/70/100/130 keV) were reconstructed with soft (Qr40) and sharp (Qr72) kernels and quantum iterative reconstruction (QIR) levels. Findings were compared with a clinically used EID-CT protocol. HU line profiles and edge-width-at-half-maximum (EWHM) at the iodine–phantom interface were measured. Reference EWHM was 0.125 mm. Cardiac half-rotation Qr40 reconstructions yielded EWHM 0.68–0.75 mm, whereas Qr72 approached reference values (0.18–0.23 mm). In LAP5.5mm, the 1-mm LAP layer was detectable on standard and UHR images with sharp kernels and higher QIR, with best attenuation fidelity using Qr72. For the smallest LAP4mm, only Qr72/QIR2 resolved the 1-mm LAP component, although measured attenuation remained elevated because of blooming from adjacent iodine and calcium. In LAP7mm, Qr72/QIR4 most closely reproduced the nominal LAP attenuation. Lower-keV images (40–70) increased LAP masking from iodine. EID-CT did not clearly depict LAP4mm.
Conclusion
PCD-CT provides superior spatial resolution and HU fidelity for detecting LAP, particularly when using UHR images combined with sharp kernels (Qr72) and higher iterative strength (QIR 2–4). LAP detection improved at 130 keV vs. lower energies.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Heart and Vascular Center, Semmelweis University, Városmajor Street 68, Budapest 1122, Hungary.
- Gottsegen National Cardiovascular Center, Budapest, Hungary.
- Physiological Controls Research Center, University Research and Innovation Center, Óbuda University, Budapest, Hungary.
- Unit of Cardiovascular Imaging (UNICA), Hospital da Luz, Lisbon, Portugal.
- Medical Imaging Centre, Semmelweis University, Budapest, Hungary.
- Scripps Clinic, Scripps Health, La Jolla, CA, USA.
ORCID-Profile: Bálint Szilveszter, Márton Kolossváry, Milán Vecsey-Nagy, Melinda Boussoussou, Pál Maurovich-Horvat, Elliot McVeigh
Förderung
- European Union
- János Bolyai Research Scholarship
- STARTING GRANT
- Hungarian National Research Development and Innovation Office
- Siemens Healthineers
- New National Excellence Program
- Ministry for Innovation and Technology
- Hungarian Academy of Sciences
Zitieren
APA
Szilveszter, B., Kolossváry, M., Kubovje, A., Marques, H., Vattay, B., Jokkel, Z., Sipos, B., Vecsey-Nagy, M., Boussoussou, M., Maurovich-Horvat, P., Newlander, S., Wesbey, G., & McVeigh, E. (2026). Mixed coronary plaque phantom analysis by photon-counting CT: impact of calcium and iodine on low-attenuation plaque detection. European Heart Journal - Imaging Methods and Practice, 4(3). https://doi.org/10.1093/ehjimp/qyag119
Vancouver
Szilveszter B, Kolossváry M, Kubovje A, Marques H, Vattay B, Jokkel Z, et al. Mixed coronary plaque phantom analysis by photon-counting CT: impact of calcium and iodine on low-attenuation plaque detection. European Heart Journal - Imaging Methods and Practice. 2026;4(3). doi:10.1093/ehjimp/qyag119
BibTeX
@article{Szilveszter2026,
author = {Szilveszter, Bálint and Kolossváry, Márton and Kubovje, Anikó and Marques, Hugo and Vattay, Borbála and Jokkel, Zsófia and Sipos, Barbara and Vecsey-Nagy, Milán and Boussoussou, Melinda and Maurovich-Horvat, Pál and Newlander, Shawn and Wesbey, George and McVeigh, Elliot},
title = {Mixed coronary plaque phantom analysis by photon-counting CT: impact of calcium and iodine on low-attenuation plaque detection},
journal = {European Heart Journal - Imaging Methods and Practice},
year = {2026},
volume = {4},
number = {3},
doi = {10.1093/ehjimp/qyag119}
}
EUVIMED