Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy
European Heart Journal - Imaging Methods and Practice · 25.07.2024 · 2 (3)
DOI: 10.1093/ehjimp/qyae082 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Heart Journal - Imaging Methods and Practice, online erschienen am 29.08.2024. Verfasst von Linus Ohlsson und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiothoracic and Vascular Surgery, Linköping University, 581 83 Linköping, Sweden. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swedish Heart-Lung Foundation, Swedish Research Council, ALF Grants Region Östergötland.
Abstract (Originalsprache)
Aims
The HeartMate 3 (HM3) implantable left ventricular assist device connects the left ventricle apex to the aorta via an outflow graft. Extrinsic obstruction of the graft (eOGO) is associated with serious morbidity and mortality and recently led to a Food and Drug Administration Class 1 device recall of HM3. This study aimed to provide a better understanding of the haemodynamic impact of extrinsic stenoses. Methods and results Computed tomography (CT) images of two retrospectively identified patients with eOGO (29 and 36% decrease in cross-sectional area, respectively, by radiological evaluation) were acquired with a novel photon-counting CT system. Numerical evaluations of haemodynamics were conducted using a high-fidelity 3D computational fluid dynamics approach on both the patient-specific graft geometries and in two virtually augmented stenotic severities and three device flows. Visual analysis identified increased velocity, pressure, and turbulent flow in the outer anterior curvature of the outflow graft; however, changes in graft pressure gradients were slight (1–9 mmHg) across the range of stenosis severities and flow rates tested.
Conclusion
Evidence of eOGO during HM3 support and the recent device recall can provoke clinical apprehension and interventions. The haemodynamic impact of a stenosis detected visually or by quantification of cross-sectional area reduction may be difficult to predict and easily overestimated. This numerical study suggests that, for clinically encountered flow rates and stenosis severities below 61% in cross-sectional area decrease, eOGO may have low haemodynamic impact. This suggests that patients without symptoms or signs consistent with haemodynamically significant obstruction might be managed expectantly.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiothoracic and Vascular Surgery, Linköping University, 581 83 Linköping, Sweden.
- Department of Health, Medicine and Caring Sciences, Linköping University, 581 83 Linköping, Sweden.
- Center of Medical Image Science and Visualization (CMIV), Linköping University, 581 83 Linköping, Sweden.
- Department of Radiology in Linköping, Linköping University, Linköping, Sweden.
- Department of Cardiology in Linköping, Linköping University, Linköping, Sweden.
- Department of Medicine, University of California San Francisco, San Francisco, CA, USA.
ORCID-Profile: Linus Ohlsson, Tino Ebbers
Förderung
- Swedish Heart-Lung Foundation
- Swedish Research Council
- ALF Grants Region Östergötland
- Sweden’s Innovation Agency Vinnova
- Foundation Linköpings Läkarsällskap
Zitieren
APA
Ohlsson, L., Sandstedt, M., Papageorgiou, J. M., Svensson, A., Bolger, A., Tamás, É., Granfeldt, H., Ebbers, T., & Lantz, J. (2024). Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy. European Heart Journal - Imaging Methods and Practice, 2(3). https://doi.org/10.1093/ehjimp/qyae082
Vancouver
Ohlsson L, Sandstedt M, Papageorgiou JM, Svensson A, Bolger A, Tamás É, et al. Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy. European Heart Journal - Imaging Methods and Practice. 2024;2(3). doi:10.1093/ehjimp/qyae082
BibTeX
@article{Ohlsson2024,
author = {Ohlsson, Linus and Sandstedt, Mårten and Papageorgiou, Joanna-Maria and Svensson, Anders and Bolger, Ann and Tamás, Éva and Granfeldt, Hans and Ebbers, Tino and Lantz, Jonas},
title = {Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy},
journal = {European Heart Journal - Imaging Methods and Practice},
year = {2024},
volume = {2},
number = {3},
doi = {10.1093/ehjimp/qyae082}
}