Assessment of photoplethysmography-based blood pressure determinations during long-term and short-term remote cardiac monitoring: the RECAMO study

Mariska van Vliet, Jan J J Aalberts, Cora Hamelinck, Arnaud D Hauer, Dieke Hoftijzer, Stefan H J Monnink, Jurjan C Schipper, Jan C Constandse, Nicholas S Peters iD, Gregory Y H Lip, Steven R Steinhubl, Eelko Ronner iD

European Heart Journal - Digital Health · 22.07.2025 · 6 (4) · 763-771

DOI: 10.1093/ehjdh/ztaf027 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Heart Journal - Digital Health, online erschienen am 27.03.2025. Verfasst von Mariska van Vliet und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Reinier the Graaf Hospital, Reinier de Graafweg 5, 2625 AD Delft, The Netherlands. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Corsano Health B.V..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Cardiovascular diseases are a global health crisis, with hypertension as a significant risk factor. Traditional cuff-based blood pressure measurements have various limitations, prompting the exploration of photoplethysmography as an alternative for continuous monitoring. This study aimed to assess a cuff-calibrated wrist-worn photoplethysmography-based blood pressure device against European Society of Hypertension recommendations. Methods and results The study assessed photoplethysmography-based blood pressure measurement stability over 28 days in 150 patients by comparing measurements of the wrist-worn photoplethysmography-based device against three daily automated reference blood pressure measurements. Additionally, awake–asleep blood pressure changes were analysed in 40 patients receiving 24-h ambulatory blood pressure monitoring. Data analysis included overall accuracy and recalibration needs during long-term monitoring, the accuracy of monitoring awake–asleep blood pressure changes, and resilience against hydrostatic pressure changes due to variations in device position. Across 28 days, mean errors of 3.84 mmHg (SD 4.46) for systolic and 4.08 mmHg (SD 3.97) for diastolic blood pressure were achieved. Before recalibration on Day 28, mean errors were 2.49 (SD 3.10) for systolic and 2.98 (SD 3.48) for diastolic blood pressure. Awake–asleep blood pressure change accuracy was demonstrated with mean errors of 2.36 (SD ± 2.40) for systolic and 2.17 (SD ± 2.13) for diastolic blood pressure. Hydrostatic pressure testing indicated resilience against changes in device position.

Conclusion

The studied wrist-worn photoplethysmography-based device demonstrated accurate and stable blood pressure monitoring over 28 days, during awake–asleep blood pressure changes and hydrostatic pressure changes. These findings support the device’s potential for remote patient monitoring. Study registration ClinicalTrials.gov identifier: NCT05899959.

Schlagwörter

Stichwörter der Autor:innen

Ambulatory blood pressure monitoringBlood pressureLong-term evaluationPhotoplethysmographyRemote monitoringWearable diagnostics

Autor:innen und Institutionen

  • Department of Cardiology, Reinier the Graaf Hospital, Reinier de Graafweg 5, 2625 AD Delft, The Netherlands.
  • National Heart and Lung Institute, Imperial College London, London, UK.
  • Department of Cardiac Electrophysiology, Imperial College Healthcare NHS Trust, London, UK.
  • Liverpool Centre for Cardiovascular Science at the University of Liverpool, Liverpool John Moores University, Liverpool Heart and Chest Hospital, Liverpool, UK.
  • Danish Centre for Health Services Research, Department of Clinical Medicine, Aalborg University, Aalborg, Denmark.
  • Weldon School of Biomedical Engineering, Purdue University, West Lafayette, IN, USA.

ORCID-Profile: Nicholas S Peters, Eelko Ronner

Förderung

  • Corsano Health B.V.

Zitieren

APA

van Vliet, M., Aalberts, J. J. J., Hamelinck, C., Hauer, A. D., Hoftijzer, D., Monnink, S. H. J., Schipper, J. C., Constandse, J. C., Peters, N. S., Lip, G. Y. H., Steinhubl, S. R., & Ronner, E. (2025). Assessment of photoplethysmography-based blood pressure determinations during long-term and short-term remote cardiac monitoring: the RECAMO study. European Heart Journal - Digital Health, 6(4), 763-771. https://doi.org/10.1093/ehjdh/ztaf027

Vancouver

van Vliet M, Aalberts JJJ, Hamelinck C, Hauer AD, Hoftijzer D, Monnink SHJ, et al. Assessment of photoplethysmography-based blood pressure determinations during long-term and short-term remote cardiac monitoring: the RECAMO study. European Heart Journal - Digital Health. 2025;6(4):763-771. doi:10.1093/ehjdh/ztaf027

BibTeX
@article{vanVliet2025,
  author = {van Vliet, Mariska and Aalberts, Jan J J and Hamelinck, Cora and Hauer, Arnaud D and Hoftijzer, Dieke and Monnink, Stefan H J and Schipper, Jurjan C and Constandse, Jan C and Peters, Nicholas S and Lip, Gregory Y H and Steinhubl, Steven R and Ronner, Eelko},
  title = {Assessment of photoplethysmography-based blood pressure determinations during long-term and short-term remote cardiac monitoring: the RECAMO study},
  journal = {European Heart Journal - Digital Health},
  year = {2025},
  volume = {6},
  number = {4},
  pages = {763--771},
  doi = {10.1093/ehjdh/ztaf027}
}
Kostenlos starten →