Background
Rotational atherectomy is crucial for heavily calcified coronary lesions, but complications like burr entrapment, though rare, can be fatal. Case Summary A 55-year-old man with diffuse coronary stenosis underwent rotational atherectomy for a left circumflex artery lesion. During the procedure, the rotational atherectomy burr disconnected from the guidewire, and the wire subsequently disengaged entirely from the burr, leaving the burr retained. A novel wire-entanglement technique was successfully employed by advancing a Fielder XT-A guidewire through the burr's lumen and intentionally knuckling a Sion guidewire to intertwine with it. This maneuver loosened and dislodged the burr, allowing for controlled proximal migration and retrieval into the guiding catheter.
Discussion
This case presents a rare scenario of complete rotational atherectomy guidewire disengagement from the burr. The successful retrieval using a wire-entanglement technique offers a valuable percutaneous bailout strategy for similar challenging situations where traditional retrieval methods may fail.
Zitieren
APA
Ding, Z., Wu, M., Qin, Q., Xu, R., & Ma, J. (2026). Wire-entanglement retrieval of a detached rotational atherectomy burr after complete wire disengagement: a case report. European Heart Journal - Case Reports. https://doi.org/10.1093/ehjcr/ytag830
Vancouver
Ding Z, Wu M, Qin Q, Xu R, Ma J. Wire-entanglement retrieval of a detached rotational atherectomy burr after complete wire disengagement: a case report. European Heart Journal - Case Reports. 2026. doi:10.1093/ehjcr/ytag830
BibTeX
@article{Ding2026,
author = {Ding, Zhenzhen and Wu, Mengzhang and Qin, Qing and Xu, Rende and Ma, Jianying},
title = {Wire-entanglement retrieval of a detached rotational atherectomy burr after complete wire disengagement: a case report},
journal = {European Heart Journal - Case Reports},
year = {2026},
doi = {10.1093/ehjcr/ytag830}
}
EUVIMED