Pulsed field ablation of a focal atrial tachycardia from the superior vena cava in proximity to the phrenic nerve: a case report

Konstantinos Tampakis iD, Evangelia-Erasmia Papakonstantinou, Alexandros Sykiotis, Sokratis Pastromas, George Andrikopoulos

European Heart Journal - Case Reports · 01.02.2025 · 9 (2)

DOI: 10.1093/ehjcr/ytaf048 ↗

Kurz gefasst

Fallbericht in der Zeitschrift European Heart Journal - Case Reports, online erschienen am 31.01.2025. Verfasst von Konstantinos Tampakis und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrophysiology and Pacing, Henry Dunant Hospital Center, 107 Mesogeion ave, Athens 11526, Greece. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Arrhythmias Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Right phrenic nerve (PN) injury is a major complication of thermal ablation of atrial tachycardias (ATs) originated from the superior vena cava (SVC). Case summary We report the case of a 41-year-old female patient admitted for catheter ablation of a frequent paroxysmal AT resistant to antiarrhythmic drugs. Electroanatomical activation map demonstrated a focal origin located at the lateral aspect of the SVC, ∼17 mm above the breakthrough of the sinus node wavefront. Importantly, high-output pacing from this site resulted in PN capture. To avoid PN injury, low-output radiofrequency (RF) ablation, with a power output limited to 20 W, was performed. However, this approach was insufficient to terminate AT. High-power RF applications in proximity to the PN were avoided and pulsed-field ablation (PFA) with a pentaspline catheter was chosen. The catheter was advanced into the SVC to the level of the earliest activation under fluoroscopic guidance and visualization within the mapping system. Two pairs of applications, in basket configuration, were delivered inside the SVC, rendering AT non-inducible while sinus node function was not compromised.

Discussion

Phrenic nerve is vulnerable to injury during ablation within the SVC using thermal ablation modalities. Low-output RF ablation may be safe but less efficient. In contrast, non-thermal approaches such as PFA may be preferable to avoid damage to the collateral tissues as PN. Electroanatomical mapping may be important to avoid lesions in proximity to the sinus node.

Schlagwörter

Stichwörter der Autor:innen

Atrial tachycardiaPhrenic nerve injuryPulsed-field ablationRadiofrequency ablationSuperior vena cava

Autor:innen und Institutionen

  • Department of Electrophysiology and Pacing, Henry Dunant Hospital Center, 107 Mesogeion ave, Athens 11526, Greece.

ORCID-Profile: Konstantinos Tampakis

Förderung

  • Arrhythmias Institute

Zitieren

APA

Tampakis, K., Papakonstantinou, E. E., Sykiotis, A., Pastromas, S., & Andrikopoulos, G. (2025). Pulsed field ablation of a focal atrial tachycardia from the superior vena cava in proximity to the phrenic nerve: a case report. European Heart Journal - Case Reports, 9(2). https://doi.org/10.1093/ehjcr/ytaf048

Vancouver

Tampakis K, Papakonstantinou EE, Sykiotis A, Pastromas S, Andrikopoulos G. Pulsed field ablation of a focal atrial tachycardia from the superior vena cava in proximity to the phrenic nerve: a case report. European Heart Journal - Case Reports. 2025;9(2). doi:10.1093/ehjcr/ytaf048

BibTeX
@article{Tampakis2025,
  author = {Tampakis, Konstantinos and Papakonstantinou, Evangelia-Erasmia and Sykiotis, Alexandros and Pastromas, Sokratis and Andrikopoulos, George},
  title = {Pulsed field ablation of a focal atrial tachycardia from the superior vena cava in proximity to the phrenic nerve: a case report},
  journal = {European Heart Journal - Case Reports},
  year = {2025},
  volume = {9},
  number = {2},
  doi = {10.1093/ehjcr/ytaf048}
}
Kostenlos starten →