EUVIMEDEUVIMED entdecken
Multizentrische Studie

Reproducibility of cardiac computed tomography classifications of hypoattenuated thickening and peridevice leak following left atrial appendage closure

Anders Kramer iD, Gerardo Lo Russo iD, Hasan S Alarouri iD, Jeremy D Collins iD, Jesper Møller Jensen iD, Jens Erik Nielsen-Kudsk iD, Mohamad Alkhouli iD, Kasper Korsholm iD

European Heart Journal - Cardiovascular Imaging · 31.12.2024 · 26 (1) · 62-71

DOI: 10.1093/ehjci/jeae236 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift European Heart Journal - Cardiovascular Imaging, online erschienen am 03.09.2024. Verfasst von Anders Kramer und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark. Hauptthemen laut MEDLINE: Atrial Appendage, Tomography, X-Ray Computed, Atrial Fibrillation. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: Eva and Henry Frænkels Memorial Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

To assess the reproducibility of interpreting hypoattenuated thickening (HAT) and peridevice leak (PDL) using cardiac computed tomography (CT) imaging following Watchman FLX left atrial appendage closure (LAAC). Methods and results In this multicentre retrospective reproducibility study, 100 anonymized post-LAAC cardiac CT scans were evaluated within the same cardiac phase by an experienced and a novice rater blinded to prior evaluations. All scans were evaluated twice by each rater, assessing overall HAT and PDL categories as well as specific associated findings based on suggested algorithms for post-LAAC interpretation. Inter- and intra-rater agreement and reliability were evaluated using absolute agreement, Cohen’s kappa, and Kendall’s tau for categorical variables, and mean difference, Bland–Altman plots, limits of agreement (lower and upper), and intra-class correlation coefficients (ICCs) for continuous variables. Within overall categories of both HAT and PDL, substantial agreement (kappa ≥0.61) and reliability (Kendall’s tau-b ≥ 0.75) were observed. Specifically, identifying high-grade HAT (kappa ≥0.78) and distal patency (kappa ≥0.85) displayed the highest agreement within HAT and PDL interpretation. Meanwhile, measuring the height of the proximal screw hub cove represented the least reliable HAT assessment among both inter- and intra-rater comparisons (ICC < 0.75), while suspected leak mechanism represented the least reproducible PDL measure.

Conclusion

Despite only minimal training of one rater, overall high levels of inter- and intra-rater agreement and reliability were observed across the chosen algorithms for interpretation of HAT and PDL following Watchman FLX LAAC. The prognostic implications of the included variables are to be explored in future trials and registries.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial AppendageTomography, X-Ray ComputedAtrial FibrillationHumansRetrospective StudiesFemaleMaleReproducibility of ResultsAgedMiddle AgedCardiac Surgical ProceduresObserver VariationPostoperative ComplicationsLeft Atrial Appendage Closure

Stichwörter der Autor:innen

Watchman FLXcardiac computed tomographydevice-related thrombosisleft atrial appendage closureperidevice leak

Autor:innen und Institutionen

  • Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark.
  • Department of Clinical Medicine, Aarhus University, Aarhus, Denmark.
  • Department of Cardiovascular Medicine, Mayo Clinic, Rochester, MN, USA.
  • Department of Clinical Sciences and Community Health, Cardiovascular Section, University of Milan, Milan, Italy.
  • Department of Radiology, Mayo Clinic, Rochester, MN, USA.

ORCID-Profile: Anders Kramer, Gerardo Lo Russo, Hasan S Alarouri, Jeremy D Collins, Jesper Møller Jensen, Jens Erik Nielsen-Kudsk, Mohamad Alkhouli, Kasper Korsholm

Förderung

  • Eva and Henry Frænkels Memorial Fund

Zitieren

APA

Kramer, A., Lo Russo, G., Alarouri, H. S., Collins, J. D., Møller Jensen, J., Nielsen-Kudsk, J. E., Alkhouli, M., & Korsholm, K. (2024). Reproducibility of cardiac computed tomography classifications of hypoattenuated thickening and peridevice leak following left atrial appendage closure. European Heart Journal - Cardiovascular Imaging, 26(1), 62-71. https://doi.org/10.1093/ehjci/jeae236

Vancouver

Kramer A, Lo Russo G, Alarouri HS, Collins JD, Møller Jensen J, Nielsen-Kudsk JE, et al. Reproducibility of cardiac computed tomography classifications of hypoattenuated thickening and peridevice leak following left atrial appendage closure. European Heart Journal - Cardiovascular Imaging. 2024;26(1):62-71. doi:10.1093/ehjci/jeae236

BibTeX
@article{Kramer2024,
  author = {Kramer, Anders and Lo Russo, Gerardo and Alarouri, Hasan S and Collins, Jeremy D and Møller Jensen, Jesper and Nielsen-Kudsk, Jens Erik and Alkhouli, Mohamad and Korsholm, Kasper},
  title = {Reproducibility of cardiac computed tomography classifications of hypoattenuated thickening and peridevice leak following left atrial appendage closure},
  journal = {European Heart Journal - Cardiovascular Imaging},
  year = {2024},
  volume = {26},
  number = {1},
  pages = {62--71},
  doi = {10.1093/ehjci/jeae236}
}
Kostenlos starten →