Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2025. Verfasst von Han-Gyeol Yeom und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Radiology and Wonkwang Dental Research Institute, College of Dentistry, Wonkwang University, Iksan, 54538, Repub…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiography, Panoramic, Radiographic Magnification, Radiography, Dental, Digital. Die Arbeit nennt 22 Literaturangaben. Gefördert von: Wonkwang University in 2024.
Abstract (Originalsprache)
Objectives
This study aimed to investigate the vertical magnification ratios of various digital panoramic radiography devices and assess their consistency. It also xamined the relationship between magnification ratios and Digital Imaging and Communications in Medicine (DICOM) header information to determine the potential use of key metadata in inferring or standardizing magnification corrections.
Methods
A ball phantom with metal spheres and an implant phantom were used to assess vertical magnification ratios. Eleven panoramic radiography systems were tested, and three DICOM tags—pixel spacing, imager pixel spacing, and aspect ratio—were analyzed. Three specialists measured phantom lengths in randomized DICOM images using PACS software, repeating measurements after 4 weeks. Magnification ratios were calculated by averaging 24 measurements per sample and comparing them to actual lengths.
Results
Intra- and interobserver reliability were high (ICC > 0.99). Most devices showed magnification ratios of 0.95–1.08, whereas those of devices 3 and 9 ranged from 1.21 to 1.30. Although some magnification ratios, which were used for system-based calibration, could be indirectly inferred by dividing the imager pixel spacing value by the pixel spacing value when both values were precisely recorded, no DICOM headers explicitly provided magnification ratios.
Conclusions
This study demonstrated that many digital panoramic radiography devices apply software-based corrections to achieve a vertical magnification ratio close to 1.0; however, inconsistencies remain across different systems. Measurement accuracy and clinical reliability can be enhanced by standardization of magnification correction methods and improved documentation in product specifications and DICOM headers, which is clinically relevant for increasing the accuracy of vertical measurements in procedures such as dental implant planning and thereby enhancing treatment predictability and patient safety.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oral and Maxillofacial Radiology and Wonkwang Dental Research Institute, College of Dentistry, Wonkwang University, Iksan, 54538, Republic of Korea.
- Department of Oral and Maxillofacial Radiology, School of Dentistry and Dental Research Institute, Seoul National University, Seoul, 03080, Republic of Korea.
- Department of Periodontology, Daejeon Dental Hospital, College of Dentistry, Wonkwang University, Daejeon, 35233, Republic of Korea.
ORCID-Profile: Han-Gyeol Yeom, Hyun-Gab Kim, Yeon-Tae Kim, Kyung-Hoe Huh, Won-Jin Yi, Min-Suk Heo, Sam-Sun Lee, Jo-Eun Kim
Förderung
- Wonkwang University in 2024
Zitieren
APA
Yeom, H. G., Kim, H. G., Kim, Y. T., Huh, K. H., Yi, W. J., Heo, M. S., Lee, S. S., & Kim, J. E. (2025). Assessment of vertical magnification ratios in digital panoramic radiography using phantoms: evaluation of DICOM metadata and calibration consistency. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twaf074
Vancouver
Yeom HG, Kim HG, Kim YT, Huh KH, Yi WJ, Heo MS, et al. Assessment of vertical magnification ratios in digital panoramic radiography using phantoms: evaluation of DICOM metadata and calibration consistency. Dentomaxillofacial Radiology. 2025. doi:10.1093/dmfr/twaf074
BibTeX
@article{Yeom2025,
author = {Yeom, Han-Gyeol and Kim, Hyun-Gab and Kim, Yeon-Tae and Huh, Kyung-Hoe and Yi, Won-Jin and Heo, Min-Suk and Lee, Sam-Sun and Kim, Jo-Eun},
title = {Assessment of vertical magnification ratios in digital panoramic radiography using phantoms: evaluation of DICOM metadata and calibration consistency},
journal = {Dentomaxillofacial Radiology},
year = {2025},
doi = {10.1093/dmfr/twaf074}
}
EUVIMED