In vitro evaluation of the accuracy of electronic apex locators and cone-beam CT in the detection of oblique root fractures
Dentomaxillofacial Radiology · 2024
DOI: 10.1093/dmfr/twae037 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Dentomaxillofacial Radiology, 2024. Verfasst von Simay Koç und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Endodontics, Faculty of Dentistry, Akdeniz University, Antalya, 07070, Turkey. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Tooth Fractures, Tooth Apex, Incisor. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben.
Abstract (Originalsprache)
Objectives
This study aimed to compare the accuracy of cone-beam CT (CBCT) scanning and 3 different electronic apex locators (EALs) in the detection of simulated oblique root fractures (ORF) in different localizations.
Methods
The study utilised a total of 80 human maxillary incisors, which were categorised into two groups based on the location of the ORF (apical and middle third of the root) formed on the buccal side of the root surface. The measurement of the distance between the incisal edge and the intersection of the ORF with the root canal was conducted using a stereomicroscope. This measurement is referred to as the actual working length (AWL). Additionally, three EALs—Dentaport ZX, EndoRadar Pro, and Propex II—were utilised to determine the electronic working length (EWL). Furthermore, CBCT images were employed to assess the distance, known as the CBCT working length (CWL). The differences were determined by subtracting AWL from EWL and CWL.
Results
Based on the accuracy of the devices, there were no significant differences observed among Dentaport ZX, EndoRadar, Propex II, and CBCT measures in both the apical and middle third ORF groups, within the acceptable range of 0.5 and 1 mm.
Conclusions
The accuracy of Dentaport ZX, Propex II, and CBCT was higher in the middle third ORF group compared to the apical third ORF group, with a tolerance of 0.5 mm. However, there were no significant differences seen among the devices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Endodontics, Faculty of Dentistry, Akdeniz University, Antalya, 07070, Turkey.
- Department of Medical Biotechnology, Health Sciences Institute, Akdeniz University, Antalya, 07070, Turkey.
ORCID-Profile: Hatice Harorlı
Zitieren
APA
Koç, S., Harorlı, H., & Kuştarcı, A. (2024). In vitro evaluation of the accuracy of electronic apex locators and cone-beam CT in the detection of oblique root fractures. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twae037
Vancouver
Koç S, Harorlı H, Kuştarcı A. In vitro evaluation of the accuracy of electronic apex locators and cone-beam CT in the detection of oblique root fractures. Dentomaxillofacial Radiology. 2024. doi:10.1093/dmfr/twae037
BibTeX
@article{Ko2024,
author = {Koç, Simay and Harorlı, Hatice and Kuştarcı, Alper},
title = {In vitro evaluation of the accuracy of electronic apex locators and cone-beam CT in the detection of oblique root fractures},
journal = {Dentomaxillofacial Radiology},
year = {2024},
doi = {10.1093/dmfr/twae037}
}
EUVIMED