Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: a systematic review and meta-analysis
Dentomaxillofacial Radiology · 29.04.2024 · 53 (4) · 207-221
DOI: 10.1093/dmfr/twae007 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Dentomaxillofacial Radiology, online erschienen am 01.03.2024. Verfasst von Ana Luiza E Carneiro und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Stomatology, School of Dentistry, University of São Paulo (USP), São Paulo, 05508-000, Brazil. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Radiation Dosage, Patient Care Planning. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior-Brasil, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior—Brasil.
Abstract (Originalsprache)
Objectives
The aim of this systematic review was to verify the accuracy of linear measurements performed on low-dose CBCT protocols for implant planning, in comparison with those performed on standard and high-resolution CBCT protocols.
Methods
The literature search included four databases (Pubmed, Web of Science, Embase, and Scopus). Two reviewers independently screened titles/abstracts and full texts according to eligibility criteria, extracted the data, and examined the methodological quality. Risk of bias assessment was performed using the Quality Assessment Tool For In Vitro Studies. Random-effects meta-analysis was used for pooling measurement error data.
Results
The initial search yielded 4684 titles. In total, 13 studies were included in the systematic review, representing a total of 81 samples, while 9 studies were included in the meta-analysis. The risk of bias ranged from medium to low. The main results across the studies indicate a strong consistency in linear measurements performed on low-dose images in relation to the reference methods. The overall pooled planning measurement error from low-dose CBCT protocols was −0.24 mm (95% CI, −0.52 to 0.04) with a high level of heterogeneity, showing a tendency for underestimation of real values. Various studies found no significant differences in measurements across different protocols (eg, voxel sizes, mA settings, or dose levels), regions (incisor, premolar, molar) and types (height vs. width). Some studies, however, noted exceptions in measurements performed on the posterior mandible.
Conclusion
Low-dose CBCT protocols offer adequate precision and accuracy of linear measurements for implant planning. Nevertheless, diagnostic image quality needs must be taken into consideration when choosing a low-dose CBCT protocol.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Stomatology, School of Dentistry, University of São Paulo (USP), São Paulo, 05508-000, Brazil.
- Section for Oral Ecology and Inflammation, Department of Dentistry and Oral Health, Aarhus University, Aarhus, 8000, Denmark.
- Steno Diabetes Center Aarhus, Aarhus, 8200, Denmark.
- Department of Dentistry and Oral Health-Section for Oral Radiology and Endodontics, Aarhus University, Aarhus, 8000, Denmark.
ORCID-Profile: Ana Luiza E Carneiro
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior-Brasil
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior—Brasil
Zitieren
APA
Carneiro, A. L. E., Reis, I. N. R., Bitencourt, F. V., Salgado, D. M. R. A., Costa, C., & Spin-Neto, R. (2024). Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: a systematic review and meta-analysis. Dentomaxillofacial Radiology, 53(4), 207-221. https://doi.org/10.1093/dmfr/twae007
Vancouver
Carneiro ALE, Reis INR, Bitencourt FV, Salgado DMRA, Costa C, Spin-Neto R. Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: a systematic review and meta-analysis. Dentomaxillofacial Radiology. 2024;53(4):207-221. doi:10.1093/dmfr/twae007
BibTeX
@article{Carneiro2024,
author = {Carneiro, Ana Luiza E and Reis, Isabella N R and Bitencourt, Fernando Valentim and Salgado, Daniela M R A and Costa, Claudio and Spin-Neto, Rubens},
title = {Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: a systematic review and meta-analysis},
journal = {Dentomaxillofacial Radiology},
year = {2024},
volume = {53},
number = {4},
pages = {207--221},
doi = {10.1093/dmfr/twae007}
}
EUVIMED