The Analytical Performance of Neonatal Total Bilirubin Assays Does Not Meet the Clinical Need

Marlies Oostendorp, Eline A Van der Hagen, Stanley Lo, Vincent Delatour, Denis Grote-Koska, Antje Staaden, Marieke A Frasa, Jenny E Kootstra-Ros, Lauren Westenberg, Libor Vítek, Aleš Dvořák, David Křepelka und 3 weitere

Clinical Chemistry · 05.03.2026 · 72 (3) · 419-421

DOI: 10.1093/clinchem/hvaf175 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Chemistry, online erschienen am 12.12.2025. Verfasst von Marlies Oostendorp und 14 weiteren Autor:innen. Beteiligte Einrichtung: Dicoon BV, Department of Clinical Chemistry and Immunology, Location Rijnstate Hospital, Arnhem, The Netherlands. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 5 Literaturangaben. Gefördert von: Siemens Healthineers, QuidelOrtho, Quality Foundation of the Dutch Medical Specialists.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Severe neonatal hyperbilirubinemia is a critical condition that requires phototherapy (PT) and/or exchange transfusion to prevent irreversible neurological sequelae. Management of neonatal hyperbilirubinemia is largely based on total bilirubin (TBIL) concentration. However, TBIL measurements are known for interassay differences, possibly leading to discrepant treatment decisions depending on the analytical method (1). Past studies using adult serum may not properly reflect the analytical performance of neonatal samples as between-method differences vary between the adult and neonatal matrix (1). The aim of this study was to compare the performance of 10 commonly used bilirubin assays in neonatal samples to the reference measurement procedure (RMP) and evaluate whether bilirubin conjugates or photoisomers contribute to the between-method differences. Methods used include BuBc slide (QuidelOrtho), vanadate oxidase (Siemens), spectrophotometry (Radiometer), and 7 diazo (Abbott, 2x Beckman Coulter, 2x Siemens, 2x Roche). Residual plasma from neonates with no significant interfering hemolysis was collected before, during, or after PT. Samples from neonates with different gestational age, time after birth, and risk factors were included and were pooled according to the PT status and similar TBIL concentration range. Additionally, leftover exchange transfusion serum was collected. Altogether, a total of 11 samples in the concentration range 8.5 to 20.3 mg/dL (146–347 µmol/L) were measured in duplicate using 10 routine TBIL assays and 2 RMPs, i.e., the Doumas and Klauke methods (2, 3).

Autor:innen und Institutionen

  • Dicoon BV, Department of Clinical Chemistry and Immunology, Location Rijnstate Hospital, Arnhem, The Netherlands.
  • Dutch Foundation for Quality Assessment in Medical Laboratories (SKML), Nijmegen, The Netherlands.
  • Radboudumc, Department of Laboratory Medicine, Nijmegen, The Netherlands.
  • Department of Pathology, Medical College of Wisconsin, Milwaukee, WI, United States.
  • Laboratoire national de métrologie et d'essais, Scientific and Industrial Metrology Division, Paris, France.
  • Medizinische Hochschule Hannover, Institute of Clinical Chemistry, Hannover, Germany.

Förderung

  • Siemens Healthineers
  • QuidelOrtho
  • Quality Foundation of the Dutch Medical Specialists
  • Czech Ministry of Health
  • Programme EXCELES
  • European Union-Next Generation EU

Zitieren

APA

Oostendorp, M., Van der Hagen, E. A., Lo, S., Delatour, V., Grote-Koska, D., Staaden, A., Frasa, M. A., Kootstra-Ros, J. E., Westenberg, L., Vítek, L., Dvořák, A., Křepelka, D., Weykamp, C., Hulzebos, C. V., & Van Berkel, M. (2026). The Analytical Performance of Neonatal Total Bilirubin Assays Does Not Meet the Clinical Need. Clinical Chemistry, 72(3), 419-421. https://doi.org/10.1093/clinchem/hvaf175

Vancouver

Oostendorp M, Van der Hagen EA, Lo S, Delatour V, Grote-Koska D, Staaden A, et al. The Analytical Performance of Neonatal Total Bilirubin Assays Does Not Meet the Clinical Need. Clinical Chemistry. 2026;72(3):419-421. doi:10.1093/clinchem/hvaf175

BibTeX
@article{Oostendorp2026,
  author = {Oostendorp, Marlies and Van der Hagen, Eline A and Lo, Stanley and Delatour, Vincent and Grote-Koska, Denis and Staaden, Antje and Frasa, Marieke A and Kootstra-Ros, Jenny E and Westenberg, Lauren and Vítek, Libor and Dvořák, Aleš and Křepelka, David and Weykamp, Cas and Hulzebos, Christian V and Van Berkel, Miranda},
  title = {The Analytical Performance of Neonatal Total Bilirubin Assays Does Not Meet the Clinical Need},
  journal = {Clinical Chemistry},
  year = {2026},
  volume = {72},
  number = {3},
  pages = {419--421},
  doi = {10.1093/clinchem/hvaf175}
}
Kostenlos starten →