Extracorporeal myoglobin elimination using the adsorber CytoSorb or the high-flux HF80 dialyzer for patients with severe rhabdomyolysis: a comparative study

Justa Friebus-Kardash, Jasmin Omar, Michael Jahn, Christina Scharf, Kristina Schönfelder, Anja Gaeckler, Kristina Boss, Bartosz Tyczynski, Andreas Kribben iD

Clinical Kidney Journal · 01.05.2025 · 18 (5)

DOI: 10.1093/ckj/sfaf071 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Kidney Journal, online erschienen am 01.05.2025. Verfasst von Justa Friebus-Kardash und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nephrology, University of Duisburg-Essen, University Hospital Essen, Essen, Germany. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Critically ill patients are frequently affected by severe rhabdomyolysis, consequently requiring renal replacement therapy (RRT). We asked whether CytoSorb is more potent than continuous veno-venous hemodiafiltration (CVVHDF) with the high-flux HF80 dialyzer in removing myoglobin and improving outcome for patients with rhabdomyolysis accompanied by acute kidney injury (AKI).

Methods

The historical HF80 cohort consisted of 97 patients who underwent CVVHDF with the HF80 dialyzer between 2010 and 2016. The CytoSorb cohort consisted of 102 patients who underwent treatment with CytoSorb in combination with CVVHDF with the standard high-flux F60S dialyzer between 2018 and 2023.

Results

Patients treated with the HF80 dialyzer achieved a median relative reduction of myoglobin of 39% after 24 h of treatment and of 69% after 3 days of treatment; this reduction was similar to the median relative myoglobin elimination achieved for the CytoSorb group (40% after 24 h and 60% after 3 days). The proportions of patients in whom return of diuresis occurred were comparable between the two extracorporeal treatments. The groups did not differ significantly in in-hospital mortality rates. The decrease in the Sequential Organ Failure Assessment (SOFA) and Simplified Acute Physiology Score II (SAPS II) scores at the end of therapy was more pronounced in the CytoSorb group (SOFA P < .0001; SAPS II P < .0001) than in the HF80 group (SOFA P = .004; SAPS II P = .03). The frequency of new-onset end-stage kidney disease among survivors was similar in both groups.

Conclusions

Our analysis showed that using CytoSorb resulted in myoglobin reduction rates and clinical outcomes similar to those achieved with CVVHDF with the high-flux HF80 dialyzer for patients with rhabdomyolysis and AKI.

Schlagwörter

Stichwörter der Autor:innen

CytoSorbacute kidney injury (AKI)continuous veno-venous hemodiafiltration (CVVHDF)high-flux HF80 dialyzerrhabdomyolysis

Autor:innen und Institutionen

  • Department of Nephrology, University of Duisburg-Essen, University Hospital Essen, Essen, Germany.
  • Department of Anesthesiology, LMU University Hospital, LMU Munich, Munich, Germany.

ORCID-Profile: Andreas Kribben

Zitieren

APA

Friebus-Kardash, J., Omar, J., Jahn, M., Scharf, C., Schönfelder, K., Gaeckler, A., Boss, K., Tyczynski, B., & Kribben, A. (2025). Extracorporeal myoglobin elimination using the adsorber CytoSorb or the high-flux HF80 dialyzer for patients with severe rhabdomyolysis: a comparative study. Clinical Kidney Journal, 18(5). https://doi.org/10.1093/ckj/sfaf071

Vancouver

Friebus-Kardash J, Omar J, Jahn M, Scharf C, Schönfelder K, Gaeckler A, et al. Extracorporeal myoglobin elimination using the adsorber CytoSorb or the high-flux HF80 dialyzer for patients with severe rhabdomyolysis: a comparative study. Clinical Kidney Journal. 2025;18(5). doi:10.1093/ckj/sfaf071

BibTeX
@article{FriebusKardash2025,
  author = {Friebus-Kardash, Justa and Omar, Jasmin and Jahn, Michael and Scharf, Christina and Schönfelder, Kristina and Gaeckler, Anja and Boss, Kristina and Tyczynski, Bartosz and Kribben, Andreas},
  title = {Extracorporeal myoglobin elimination using the adsorber CytoSorb or the high-flux HF80 dialyzer for patients with severe rhabdomyolysis: a comparative study},
  journal = {Clinical Kidney Journal},
  year = {2025},
  volume = {18},
  number = {5},
  doi = {10.1093/ckj/sfaf071}
}
Kostenlos starten →