A Rapid and Validated Reversed-Phase High-Performance Liquid Chromatographic Method with Pre-Column Derivatization and Fluorescence Detection for Quantification of Twenty Amino Acids in Total Parenteral Nutrition Solutions

Duygu Konuklu, Cemil Can Eylem, Fatma Göçerler, İncilay Süslü

Journal of Chromatographic Science · 2026

DOI: 10.1093/chromsci/bmaf067 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Chromatographic Science, 2026. Verfasst von Duygu Konuklu und 3 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Pharmacy, Department of Analytical Chemistry, Hacettepe University, 06100, Ankara, Turkey. Hauptthemen laut MEDLINE: Amino Acids, Chromatography, Reverse-Phase, Parenteral Nutrition Solutions. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Accurate quantification of all 20 proteinogenic amino acids (AAs) is critical for the quality control of total parenteral nutrition (TPN) formulations containing complex excipient matrices. This study presents a rapid and validated reversed-phase high-performance liquid chromatography (RP-HPLC) method employing pre-column derivatization with 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AQC) and fluorescence detection (FLD) for the simultaneous determination of AAs. Complete baseline separation of all 20 AAs was achieved within 18 minutes, with a total analytical run time of 25 minutes on a conventional HPLC system. The method was validated in accordance with ICH Q2(R2) guidelines, demonstrating excellent linearity (R2 > 0.999), high sensitivity (limits of detection = 0.02–0.16 μmol/L), precision (relative standard deviation< 2%), accuracy (recoveries = 98.2–101.7%), specificity (no matrix interference) and robustness under deliberate minor variations in chromatographic conditions. These findings confirm the method’s reliability and suitability for pharmaceutical quality control. This RP-HPLC–FLD method provides a rapid, cost-effective and reproducible alternative to LC–MS for routine quality control of TPN formulations in pharmaceutical laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Amino AcidsChromatography, Reverse-PhaseParenteral Nutrition SolutionsChromatography, High Pressure LiquidReproducibility of ResultsLimit of DetectionSpectrometry, FluorescenceCarbamatesLinear ModelsAminoquinolines

Substanzen

Amino Acids6-aminoquinolyl-N-hydroxysuccinimidyl carbamateCarbamatesParenteral Nutrition SolutionsAminoquinolines

Autor:innen und Institutionen

  • Faculty of Pharmacy, Department of Analytical Chemistry, Hacettepe University, 06100, Ankara, Turkey.
  • Research and Development Department, Turktıpsan Pharmaceuticals, 06750, Ankara, Turkey.

Zitieren

APA

Konuklu, D., Eylem, C. C., Göçerler, F., & Süslü, İ. (2026). A Rapid and Validated Reversed-Phase High-Performance Liquid Chromatographic Method with Pre-Column Derivatization and Fluorescence Detection for Quantification of Twenty Amino Acids in Total Parenteral Nutrition Solutions. Journal of Chromatographic Science. https://doi.org/10.1093/chromsci/bmaf067

Vancouver

Konuklu D, Eylem CC, Göçerler F, Süslü İ. A Rapid and Validated Reversed-Phase High-Performance Liquid Chromatographic Method with Pre-Column Derivatization and Fluorescence Detection for Quantification of Twenty Amino Acids in Total Parenteral Nutrition Solutions. Journal of Chromatographic Science. 2026. doi:10.1093/chromsci/bmaf067

BibTeX
@article{Konuklu2026,
  author = {Konuklu, Duygu and Eylem, Cemil Can and Göçerler, Fatma and Süslü, İncilay},
  title = {A Rapid and Validated Reversed-Phase High-Performance Liquid Chromatographic Method with Pre-Column Derivatization and Fluorescence Detection for Quantification of Twenty Amino Acids in Total Parenteral Nutrition Solutions},
  journal = {Journal of Chromatographic Science},
  year = {2026},
  doi = {10.1093/chromsci/bmaf067}
}
Kostenlos starten →