EUVIMEDEUVIMED entdecken
Crossref

Design and characterization of an LC-based resonator complementary metal-oxide-semiconductor circuitry for noncontact complex dielectric constant sensor system

Chii-Wann Lin, Sheng-Yu Peng, Ya-Fu Liou, Hua-An Shih, Wei-Liang Hsu, Sheng-Hsun Huang, Dan-Jing Chang, Yun-Da Sung

Bulletin of the Chemical Society of Japan · 03.07.2025 · 98 (7)

DOI: 10.1093/bulcsj/uoaf064 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bulletin of the Chemical Society of Japan, online erschienen am 01.07.2025. Verfasst von Chii-Wann Lin und 7 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: MOST, NSTC.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The complex dielectric constant of biomaterials is rarely measured using an integrated circuit in the microwave region. Our study proposes a novel sensor system for measuring complex dielectric constants, featuring an integrated complementary metal-oxide-semiconductor (CMOS) sensing interface and readout circuits. This system comprises a designated LC oscillator operating at 1 GHz, with the material under test coupled through an interdigitated electrode antenna. The system also includes circuits for frequency and amplitude readouts, designed to detect changes in resonant frequency and amplitude loss. These changes pertain to the real and imaginary components of the sample's relative permittivity. A prototype chip was fabricated using a 0.18 μm CMOS process. The preliminary measurement results demonstrate that the fabricated sensor system successfully measures air, deionized water, and various concentrations of ethanol and methanol compared with a refractometer and a Vector Network Analyzer, with excellent linearity and an average standard deviation of frequency readout of 0.002% and an amplitude loss of approximately 0.2%, respectively. The dynamic responses of our sensor system were validated by alternating between air and deionized water samples using a syringe pump at a flow rate of up to 500 μL/min. This proposed sensor offers a promising, noninvasive, label-free, real-time sensing solution for continuous bio-manufacturing.

Förderung

  • MOST
  • NSTC

Zitieren

APA

Lin, C., Peng, S., Liou, Y., Shih, H., Hsu, W., Huang, S., Chang, D., & Sung, Y. (2025). Design and characterization of an LC-based resonator complementary metal-oxide-semiconductor circuitry for noncontact complex dielectric constant sensor system. Bulletin of the Chemical Society of Japan, 98(7). https://doi.org/10.1093/bulcsj/uoaf064

Vancouver

Lin C, Peng S, Liou Y, Shih H, Hsu W, Huang S, et al. Design and characterization of an LC-based resonator complementary metal-oxide-semiconductor circuitry for noncontact complex dielectric constant sensor system. Bulletin of the Chemical Society of Japan. 2025;98(7). doi:10.1093/bulcsj/uoaf064

BibTeX
@article{Lin2025,
  author = {Lin, Chii-Wann and Peng, Sheng-Yu and Liou, Ya-Fu and Shih, Hua-An and Hsu, Wei-Liang and Huang, Sheng-Hsun and Chang, Dan-Jing and Sung, Yun-Da},
  title = {Design and characterization of an LC-based resonator complementary metal-oxide-semiconductor circuitry for noncontact complex dielectric constant sensor system},
  journal = {Bulletin of the Chemical Society of Japan},
  year = {2025},
  volume = {98},
  number = {7},
  doi = {10.1093/bulcsj/uoaf064}
}
Kostenlos starten →