Accuracy of point-of-care testing devices for haemoglobin in the operating room: meta-analysis
BJS Open · 03.01.2024 · 8 (1)
DOI: 10.1093/bjsopen/zrad148 ↗Kurz gefasst
Metaanalyse in der Zeitschrift BJS Open, online erschienen am 24.01.2024. Verfasst von Hilalion San Ahn und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada. Hauptthemen laut MEDLINE: Hemoglobins, Operating Rooms. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 73 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Point-of-care tests (POCT) for haemoglobin are increasingly used to guide intraoperative transfusion. However, their accuracy compared to central laboratory tests is unknown. The objective was to perform a systematic review and meta-analysis of method comparison studies assessing the accuracy of POCT versus central laboratory haemoglobin tests in patients undergoing surgery.
Methods
Electronic databases were searched from inception to April 2020 (updated August 2023). Any methodological approach comparing haemoglobin measurements between POCT and central laboratory in patients undergoing surgery under anaesthesia in the operating room were included. Data abstraction was guided by PRISMA and risk of bias was assessed by QUADAS-2. Data were extracted independently and in duplicate by two reviewers. Outcomes included mean differences between POCT and central laboratory haemoglobin with associated standard deviations and 95% limits of agreement (LOA).
Results
Of 3057 citations, 34 studies were included (n = 2427, 6857 paired measurements). Several devices were compared (pulse co-oximetry, n = 25; HemoCue, n = 10; iSTAT, n = 6; blood gas analysers, n = 10; haematology analyser, n = 2). Median sample size was 41 patients, and 11 studies were funded by device manufacturers. Fifteen of 34 studies had low risk of bias. Pooled mean differences (95% LOA) were: pulse co-oximeters 2.3 g/l (−25.2–29.8), HemoCue −0.3 g/l (−11.1–10.5), iSTAT −0.3 g/l (−8.4–7.8) and blood gas analysers −2.6 g/l (−17.8–12.7).
Conclusion
All POCT examining intraoperative haemoglobin measurement yielded pooled mean difference LOAs larger than the allowable limit difference of ±4 g/dl. Intraoperative haemoglobin measured by POCT should not be considered interchangeable with central laboratory values and caution is necessary when using these tests to guide intraoperative transfusion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada.
- Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON, Canada.
- Department of Pathology and Laboratory Medicine, University of Ottawa, Ottawa, ON, Canada.
- Department of Anesthesiology & Pain Medicine, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada.
ORCID-Profile: Julie L V Shaw, Daniel I McIsaac, Guillaume Martel
Zitieren
APA
Ahn, H. S., Lenet, T., Gilbert, R. W. D., Mallick, R., Shaw, J. L. V., Fergusson, D. A., McIsaac, D. I., & Martel, G. (2024). Accuracy of point-of-care testing devices for haemoglobin in the operating room: meta-analysis. BJS Open, 8(1). https://doi.org/10.1093/bjsopen/zrad148
Vancouver
Ahn HS, Lenet T, Gilbert RWD, Mallick R, Shaw JLV, Fergusson DA, et al. Accuracy of point-of-care testing devices for haemoglobin in the operating room: meta-analysis. BJS Open. 2024;8(1). doi:10.1093/bjsopen/zrad148
BibTeX
@article{Ahn2024,
author = {Ahn, Hilalion San and Lenet, Tori and Gilbert, Richard W D and Mallick, Ranjeeta and Shaw, Julie L V and Fergusson, Dean A and McIsaac, Daniel I and Martel, Guillaume},
title = {Accuracy of point-of-care testing devices for haemoglobin in the operating room: meta-analysis},
journal = {BJS Open},
year = {2024},
volume = {8},
number = {1},
doi = {10.1093/bjsopen/zrad148}
}
EUVIMED