TPT 7.11 Comparative Evaluation of Silicone and Urethane Elastomers for Ultrasound-Compatible Vascular Phantom Construction
British Journal of Surgery · 05.08.2026 · 113 (Supplement_8)
DOI: 10.1093/bjs/znag087.250 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Surgery, online erschienen am 05.08.2026. Verfasst von Premjithlal Bhaskaran und 3 weiteren Autor:innen.
Abstract (Originalsprache)
Introduction
Vascular phantoms are indispensable for ultrasound imaging research, procedural training, and device validation. The choice of elastomeric material influences mechanical durability, acoustic fidelity, and anatomical realism. Despite widespread use of silicone and urethane, direct comparisons for vascular phantom applications remain limited.
Methods
Platsil Gel-10 and Transil 40-1 (silicone) and urethane rubber were prepared with additives to adjust hardness, transparency, and acoustic scattering. Mechanical characterization included Shore hardness, elasticity, tear resistance, and durability under repeated probe compression. Acoustic performance was assessed using 5–10 MHz ultrasound, measuring backscatter, attenuation, and vessel visualization. Multilayer phantoms replicating skin, fat, muscle, and vessels were constructed to evaluate imaging performance and Doppler signal reproducibility.
Results
Silicone elastomers offered tunable compliance, optical transparency, and fine-grained speckle patterns, closely mimicking soft tissue echotexture. Urethane exhibited superior mechanical durability, puncture resistance, and elastic recovery, maintaining structural integrity after >100 probe compressions. Both materials allowed clear vessel lumen visualization and Doppler flow measurement, though urethane provided more stable signals under pulsatile conditions. Hybrid phantoms combining silicone for superficial layers and urethane for vascular walls achieved optimal balance between mechanical realism, acoustic fidelity, and long-term stability.
Conclusion
Silicone-based phantoms are well-suited for research requiring high acoustic fidelity and tunable properties, whereas urethane is ideal for repeated training and long-term use. Hybrid constructions can leverage the advantages of both materials, enabling durable, high-fidelity vascular phantoms for ultrasound simulation, procedural training, and device evaluation.
Zitieren
APA
Bhaskaran, P., Bhaskaran, I. P., Shun-Shin, M., & Aslam, M. F. (2026). TPT 7.11 Comparative Evaluation of Silicone and Urethane Elastomers for Ultrasound-Compatible Vascular Phantom Construction. British Journal of Surgery, 113(Supplement_8). https://doi.org/10.1093/bjs/znag087.250
Vancouver
Bhaskaran P, Bhaskaran IP, Shun-Shin M, Aslam MF. TPT 7.11 Comparative Evaluation of Silicone and Urethane Elastomers for Ultrasound-Compatible Vascular Phantom Construction. British Journal of Surgery. 2026;113(Supplement_8). doi:10.1093/bjs/znag087.250
BibTeX
@article{Bhaskaran2026,
author = {Bhaskaran, Premjithlal and Bhaskaran, India Premjithlal and Shun-Shin, Mathew and Aslam, Mohammed Farid},
title = {TPT 7.11 Comparative Evaluation of Silicone and Urethane Elastomers for Ultrasound-Compatible Vascular Phantom Construction},
journal = {British Journal of Surgery},
year = {2026},
volume = {113},
number = {Supplement_8},
doi = {10.1093/bjs/znag087.250}
}
EUVIMED