1266 Stabilisation of Femoral Fractures with the Bactiguard Retrograde Intramedullary Nail: Early Results with a Novel Metal-Coated Device
British Journal of Surgery · 10.07.2026 · 113 (Supplement_6)
DOI: 10.1093/bjs/znag063.032 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Surgery, online erschienen am 10.07.2026. Verfasst von S M Wakefield und 3 weiteren Autor:innen.
Abstract (Originalsprache)
Aim
To investigate the safety and early clinical results of a novel, noble metal-coated titanium intramedullary nail (IMN) for femoral fractures in a single, level-one Major Trauma Centre.
Method
Between June 2019 and January 2024, adult patients (>16 years-old) who had sustained femoral fractures stabilised with a Bactiguard-coated (Bactiguard AB, Vasagatan, Sweden) ZNN retrograde femoral IMN (Zimmer Biomet, Indiana, USA), or a non-coated IMN, were identified. Minimum follow-up time was 12-months, or until bony-union was achieved. Data collected included: patient-age, biological-sex, co-morbidities, injury-mechanism, presence of an open injury, Injury-Severity-Score (ISS), bony-union, complications, need for re-operation and follow-up time.
Result
53 patients with femoral fractures were recruited: Bactiguard-coated IMN group (n=23), and non-coated IMN (n=30). Patients in the non-coated group had a higher median Charlson Comorbidity Score (median-score 2 vs 0 in coated group; p=0.02). Both groups were otherwise well-matched. No cases of fracture-related infection (FRI) were recorded in either group. Rates of union: 91% in Bactiguard-coated group vs 83% in non-coated group (p>0.05). Time to union was shorter in the Bactiguard-coated group (108 days vs 150 days in non-coated group; p<0.05). Complication-rates were equivalent between both groups (39% in coated-group vs 37% in non-coated group; p=0.92); however, complication type varied. The majority of complications in the Bactiguard-coated group were due to irritant screws (56%). In contrast, the most common complication in the non-coated group was non-union (45%).
Conclusions
Our findings suggest that the introduction of a new-generation Bactiguard-coated retrograde femoral IMN produced safe and clinically efficacious results, with no episodes of FRI.
Zitieren
APA
Wakefield, S. M., Rodham, P. L., Kanakaris, N., & Giannoudis, P. (2026). 1266 Stabilisation of Femoral Fractures with the Bactiguard Retrograde Intramedullary Nail: Early Results with a Novel Metal-Coated Device. British Journal of Surgery, 113(Supplement_6). https://doi.org/10.1093/bjs/znag063.032
Vancouver
Wakefield SM, Rodham PL, Kanakaris N, Giannoudis P. 1266 Stabilisation of Femoral Fractures with the Bactiguard Retrograde Intramedullary Nail: Early Results with a Novel Metal-Coated Device. British Journal of Surgery. 2026;113(Supplement_6). doi:10.1093/bjs/znag063.032
BibTeX
@article{Wakefield2026,
author = {Wakefield, S M and Rodham, P L and Kanakaris, N and Giannoudis, P},
title = {1266 Stabilisation of Femoral Fractures with the Bactiguard Retrograde Intramedullary Nail: Early Results with a Novel Metal-Coated Device},
journal = {British Journal of Surgery},
year = {2026},
volume = {113},
number = {Supplement_6},
doi = {10.1093/bjs/znag063.032}
}
EUVIMED