A transparent gel pad bolus enabling direct skin recognition in surface-guided radiation therapy
British Journal of Radiology · 21.10.2026
DOI: 10.1093/bjr/tqag256 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Radiology, online erschienen am 21.10.2026. Verfasst von Shunta Hirano und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, 1-14-1, Miharadai, Minami-ku, Sakai-shi, Osaka, 590-0197, Japa…. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Society for the Promotion of Science KAKENHI.
Abstract (Originalsprache)
Objectives
To evaluate the feasibility of repurposing a transparent ultrasound gel pad as a radiotherapy bolus, we focused on dosimetric performance and compatibility with surface-guided radiation therapy (SGRT).
Methods
A gel pad bolus composed of tamarind seed gum and polyhydric alcohols was compared with a conventional gel bolus and a thermoplastic bolus. Percent depth dose was measured for 6 MV and 10 MV photon beams and 6 MeV and 9 MeV electron beams. Straight-line optical transmittance was assessed over 380-800 nm. Surface recognition accuracy was quantified using the AlignRT SGRT system on a breast phantom.
Results
The gel pad produced a measurable bolus effect for both photon and electron beams, although superficial dose enhancement was lower than that of conventional materials (85.5% vs 90.3%-91.5% at 6 MV). Deep doses beyond dmax agreed within 1% for photon beams. At 730 nm, the gel pad bolus demonstrated high transmittance (92.3%), compared with the gel (19.8%) and HM (36.6%) boluses. The SGRT recognition rate with the gel pad bolus (89.10%) was comparable to the no-bolus condition (92.93%, P = .176), whereas conventional boluses significantly reduced recognition (P < .001).
Conclusions
The transparent gel pad bolus enabled direct SGRT-based recognition of the underlying skin surface while maintaining clinically relevant bolus functionality. Advances in knowledge Optical transparency enables SGRT monitoring through a bolus, highlighting optical compatibility as an important design consideration.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, 1-14-1, Miharadai, Minami-ku, Sakai-shi, Osaka, 590-0197, Japan.
- Department of Medical Physics, Sumitomo Hospital, 5-3-20 Nakanoshima, Kita-ku, Osaka, Osaka, 530-0005 Japan.
- Department of Central Radiology, Kindai University Hospital, 1-14-1, Miharadai, Minami-ku, Sakai-shi, Osaka, 590-0197, Japan.
- Department of Radiation Oncology, Kindai University Faculty of Medicine, 1-14-1, Miharadai, Minami-ku, Sakai-shi, Osaka, 590-0197, Japan.
Förderung
- Japan Society for the Promotion of Science KAKENHI
Zitieren
APA
Hirano, S., Matsumoto, K., Otsuka, M., Miyazaki, N., Uehara, T., & Monzen, H. (2026). A transparent gel pad bolus enabling direct skin recognition in surface-guided radiation therapy. British Journal of Radiology. https://doi.org/10.1093/bjr/tqag256
Vancouver
Hirano S, Matsumoto K, Otsuka M, Miyazaki N, Uehara T, Monzen H. A transparent gel pad bolus enabling direct skin recognition in surface-guided radiation therapy. British Journal of Radiology. 2026. doi:10.1093/bjr/tqag256
BibTeX
@article{Hirano2026,
author = {Hirano, Shunta and Matsumoto, Kenji and Otsuka, Masakazu and Miyazaki, Naoto and Uehara, Takuya and Monzen, Hajime},
title = {A transparent gel pad bolus enabling direct skin recognition in surface-guided radiation therapy},
journal = {British Journal of Radiology},
year = {2026},
doi = {10.1093/bjr/tqag256}
}
EUVIMED