Comparison of 4 CBCT-IGRT geometric QA phantoms within a single institutional workflow
British Journal of Radiology · 2026
DOI: 10.1093/bjr/tqag197 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift British Journal of Radiology, 2026. Verfasst von Hyojun Park und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital, Gyeonggi-do 14353, Republic of Korea. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Quality Assurance, Health Care, Radiotherapy, Image-Guided. Die Arbeit nennt 14 Literaturangaben.
Abstract (Originalsprache)
Objectives
To compare 4 commercial cone-beam CT (CBCT) geometric quality assurance phantoms within a common workflow, focusing on positioning/repositioning accuracy and agreement between the laser-defined setup reference and the CBCT imaging isocenter.
Methods
MIMI, QUASAR Penta-Guide, ISOCube, and MZ CUBE were evaluated monthly for 5 months on a Versa HD system with XVI CBCT and a HexaPOD couch. Two observers independently completed each measurement. Three-dimensional post-correction translational residuals (Δdpos/rep) and laser-defined setup reference-CBCT imaging isocenter distances (Δdiso) were calculated. Phantom comparisons used observer-averaged values from the 5 monthly sessions. Rotational correction and CBCT image quality were assessed descriptively with MZ CUBE.
Results
Mean Δdpos/rep values ranged from 0.19 to 0.31 mm, with mean absolute observer differences of 0.05-0.12 mm. No overall difference was found among the 4 phantoms (P = .547). Mean Δdiso values ranged from 0.68 to 0.87 mm, and all measurements were below 2 mm. An overall difference was observed for Δdiso (P = .019), although no pairwise comparisons were performed. With MZ CUBE, mean post-correction translational and rotational residuals were 0.24 mm and 0.37°, respectively, and image-quality measures showed no consistent temporal trend.
Conclusions
The 4 phantoms showed similarly small post-correction translational residuals. In our setting, MZ CUBE additionally supported rotational and image-quality checks. Advances in knowledge Direct comparison within a common workflow showed similar translational performance, while the scope of additional testing depended on the phantom configuration and locally available accessories.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital, Gyeonggi-do 14353, Republic of Korea.
- Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, Republic of Korea.
- Department of Radiation Oncology, Chung-Ang University College of Medicine, Seoul 06974, Republic of Korea.
- Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul 05368, Republic of Korea.
- Biomedical Research Institute, Seoul National University Hospital, Seoul 03080, Republic of Korea.
- Department of Radiation Oncology, Seoul National University Hospital, Seoul 03080, Republic of Korea.
ORCID-Profile: Hyojun Park, Hyejo Ryu, Jin Hwa Choi, So-Yeon Park, Yoonsuk Huh, Seonghee Kang, Chang Heon Choi, Jung-In Kim, Minsoo Chun
Zitieren
APA
Park, H., Ryu, H., Choi, J. H., Park, S. Y., Huh, Y., Kang, S., Choi, C. H., Kim, J. I., & Chun, M. (2026). Comparison of 4 CBCT-IGRT geometric QA phantoms within a single institutional workflow. British Journal of Radiology. https://doi.org/10.1093/bjr/tqag197
Vancouver
Park H, Ryu H, Choi JH, Park SY, Huh Y, Kang S, et al. Comparison of 4 CBCT-IGRT geometric QA phantoms within a single institutional workflow. British Journal of Radiology. 2026. doi:10.1093/bjr/tqag197
BibTeX
@article{Park2026,
author = {Park, Hyojun and Ryu, Hyejo and Choi, Jin Hwa and Park, So-Yeon and Huh, Yoonsuk and Kang, Seonghee and Choi, Chang Heon and Kim, Jung-In and Chun, Minsoo},
title = {Comparison of 4 CBCT-IGRT geometric QA phantoms within a single institutional workflow},
journal = {British Journal of Radiology},
year = {2026},
doi = {10.1093/bjr/tqag197}
}
EUVIMED