Results of an international survey of SunCHECK Patient users: consensus and benchmark data for plan verification, pre-treatment quality assurance, and in vivo dosimetry tolerances

Aisling Haughey, Robert Biggar, Rhydian Caines, Stephen Moloney, James Earley, Paul Doolan iD

British Journal of Radiology · 2026

DOI: 10.1093/bjr/tqag166 ↗

Kurz gefasst

Fachartikel in der Zeitschrift British Journal of Radiology, 2026. Verfasst von Aisling Haughey und 5 weiteren Autor:innen. Beteiligte Einrichtung: North West Cancer Centre, WHSCT, Derry, BT47 6SB, United Kingdom. Hauptthemen laut MEDLINE: Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, In Vivo Dosimetry. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Electronic portal imaging devices (EPID) are well suited to radiotherapy in vivo dosimetry (IVD), but limited guidance remains a barrier to routine clinical use. This SunCHECK (Sun Nuclear Corporation) user survey captured real-world configurations and tolerances for EPID-based patient-specific verification, as well as other aspects of patient-related QA, to establish consensus benchmark data and inform guidance for new and existing users.

Methods

A structured survey was distributed to SunCHECK Patient users across the United Kingdom and Europe. Respondents reported current clinical practice for treatment plan verification, patient-specific quality assurance (PSQA), and IVD, including calibration methods, analysis modes, and investigation tolerances.

Results

Sixty-four responses were received from 18 countries. Linear accelerator manufacturer, treatment planning algorithm, and level of SunCHECK experience had minimal influence on applied tolerances. Point dose difference is no longer widely used as a pass-fail metric. Most centers use gamma analysis for calibration with global normalization, a 10% threshold, and a 95% pass rate. Plan verification and PSQA are almost universally assessed using gamma criteria of 3%/3 mm with a 95% passing tolerance. IVD tolerances fall into 2 main approaches: gamma criteria of 10%/5 mm for detecting gross errors, and tighter criteria of either 3%/3 mm or 5%/5 mm for identifying smaller, clinically relevant discrepancies.

Conclusions

Despite extensive configurability within SunCHECK, there is strong consistency across centers in selected settings and investigation tolerances. These benchmark data provide practical guidance for reviewing and optimizing local configurations. Advances in knowledge This work provides consensus-based guidance for configuring EPID IVD systems to reliably identify treatment delivery inconsistencies in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quality Assurance, Health CareRadiotherapy Planning, Computer-AssistedIn Vivo DosimetryHumansBenchmarkingUnited KingdomRadiotherapy DosageEuropeSurveys and QuestionnairesParticle AcceleratorsConsensusCalibrationRadiometry

Stichwörter der Autor:innen

EPIDin vivo dosimetrypatient QA

Autor:innen und Institutionen

  • North West Cancer Centre, WHSCT, Derry, BT47 6SB, United Kingdom.
  • Royal Devon University Healthcare NHS Foundation Trust, Exeter, EX2 5DW, United Kingdom.
  • University of Liverpool and The Clatterbridge Cancer Centre NHS Foundation Trust, Liverpool, L7 8YA, United Kingdom.
  • University Hospitals Dorset NHS Foundation Trust, Poole, BH15 2JB, United Kingdom.
  • Royal Surrey NHS Foundation Trust, Guildford, GU2 7XX, United Kingdom.
  • German Oncology Centre, Limassol, 4108, Cyprus.

ORCID-Profile: Paul Doolan

Zitieren

APA

Haughey, A., Biggar, R., Caines, R., Moloney, S., Earley, J., & Doolan, P. (2026). Results of an international survey of SunCHECK Patient users: consensus and benchmark data for plan verification, pre-treatment quality assurance, and in vivo dosimetry tolerances. British Journal of Radiology. https://doi.org/10.1093/bjr/tqag166

Vancouver

Haughey A, Biggar R, Caines R, Moloney S, Earley J, Doolan P. Results of an international survey of SunCHECK Patient users: consensus and benchmark data for plan verification, pre-treatment quality assurance, and in vivo dosimetry tolerances. British Journal of Radiology. 2026. doi:10.1093/bjr/tqag166

BibTeX
@article{Haughey2026,
  author = {Haughey, Aisling and Biggar, Robert and Caines, Rhydian and Moloney, Stephen and Earley, James and Doolan, Paul},
  title = {Results of an international survey of SunCHECK Patient users: consensus and benchmark data for plan verification, pre-treatment quality assurance, and in vivo dosimetry tolerances},
  journal = {British Journal of Radiology},
  year = {2026},
  doi = {10.1093/bjr/tqag166}
}
Kostenlos starten →