Reproducibility of lung density quantification across photon-counting and conventional energy-integrating CT: a comparative study
British Journal of Radiology · 01.11.2025 · 98 (1175) · 1890-1898
DOI: 10.1093/bjr/tqaf103 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift British Journal of Radiology, online erschienen am 10.05.2025. Verfasst von Saman Sotoudeh-Paima und 2 weiteren Autor:innen. Beteiligte Einrichtung: Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University School of Medicine, Durham, …. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Lung. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Objectives
To evaluate how photon-counting CT (PCCT) reproducibly quantifies lung density compared to conventional CT (energy-integrating CT or EICT) and assess the relevance of image quality.
Methods
In this retrospective analysis of a prospective study, subjects underwent acquisitions using EICT and PCCT systems (August-December 2021) on the same day. PCCT projections were reconstructed with 4 kernels and 2 voxel sizes using 2024 reconstruction software (ReconCT, Siemens). PCCT and EICT data were compared using lung density measurements: percentage of low-attenuation areas <−950 Hounsfield unit (HU) (LAA-950), 15th percentile of lung density histogram (Perc15), and mean lung density (MLD). Measurement reproducibility and its relation to image quality parameters (modulation transfer function and noise) were analyzed.
Results
Fifty-four subjects (mean age, 67.4 years ± 9.6, 30 female) were studied. PCCT images had the closest agreement to EICT under identical voxel size and similar kernels, yielding lower LAA-950 (−0.9% ± 2.7, P < 0.05), higher Perc15 (1.4 HU ± 7.9, P > 0.05), and higher MLD (2.4 HU ± 10.8, P > 0.05). At higher PCCT resolution (1024 × 1024 matrix, 0.2 mm slice thickness), the smoother quantitative kernel (Qr36) improved agreement, with differences of −0.5% ± 3.0, −0.1 HU ± 9.8, and 2.8 HU ± 11.1, for LAA-950, Perc15, and MLD (P > 0.05). Image quality parameters strongly correlated with density measurements (RPerc152=85.3%).
Conclusions
PCCT yields reproducible lung density measurements comparable to EICT, with optimal reproducibility dependent on optimized reconstruction parameters and explainable by image quality parameters. Advances in knowledge Lung density reproducibility is critical for integrating PCCT into EICT-based workflows. In our study, PCCT aligned more closely with EICT using similar kernels at identical voxel sizes, or smoother kernels at smaller voxel sizes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University School of Medicine, Durham, NC 27705, United States.
- Department of Electrical & Computer Engineering, Duke University, Durham, NC 27708, United States.
- Medical Physics Graduate Program, Duke University, Durham, NC 27705, United States.
- Department of Biomedical Engineering, Duke University, Durham, NC 27708, United States.
- Department of Physics, Duke University, Durham, NC 27708, United States.
ORCID-Profile: Saman Sotoudeh-Paima, Ehsan Samei, Ehsan Abadi
Förderung
- NIH HHS
- National Institutes of Health
Zitieren
APA
Sotoudeh-Paima, S., Samei, E., & Abadi, E. (2025). Reproducibility of lung density quantification across photon-counting and conventional energy-integrating CT: a comparative study. British Journal of Radiology, 98(1175), 1890-1898. https://doi.org/10.1093/bjr/tqaf103
Vancouver
Sotoudeh-Paima S, Samei E, Abadi E. Reproducibility of lung density quantification across photon-counting and conventional energy-integrating CT: a comparative study. British Journal of Radiology. 2025;98(1175):1890-1898. doi:10.1093/bjr/tqaf103
BibTeX
@article{SotoudehPaima2025,
author = {Sotoudeh-Paima, Saman and Samei, Ehsan and Abadi, Ehsan},
title = {Reproducibility of lung density quantification across photon-counting and conventional energy-integrating CT: a comparative study},
journal = {British Journal of Radiology},
year = {2025},
volume = {98},
number = {1175},
pages = {1890--1898},
doi = {10.1093/bjr/tqaf103}
}
EUVIMED