Underfilled tubes revisited: What blood tests can be reported on short draws?
American Journal of Clinical Pathology · 2025
DOI: 10.1093/ajcp/aqaf109 ↗Kurz gefasst
Fachartikel in der Zeitschrift American Journal of Clinical Pathology, 2025. Verfasst von Lawrence de Koning und 6 weiteren Autor:innen. Beteiligte Einrichtung: Alberta Precision Laboratories, Diagnostic and Scientific Research Centre, Calgary, AB, Canada. Hauptthemen laut MEDLINE: Blood Specimen Collection, Phlebotomy. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Objective
Underfilled blood tubes (short draws) are often collected from children or those with poor venous access. In a pilot study, we investigated which tests among a large acute care panel could be reported on short draws.
Methods
Blood was drawn in BD vacutainers (short draw: 1 mL [33%-56% fill volume] vs complete draw: 1.8-3 mL [100% fill volume]) from 12 volunteers for 3 coagulation tests, 36 chemistry tests, and the complete blood count (CBC) with differential. Tests that were strong candidates for reporting did not have statistically significant biases between short and complete draws, whereas potential candidates had statistically significant biases that were small (<25% of total allowable error and less than desirable bias from biological variation). Biases that increased or decreased across concentration ranges invalidated reporting candidacy.
Results
Two coagulation tests, 14 chemistry tests, and 15 CBC components were strong candidates for reporting. There were 9 chemistry tests and 2 CBC components that were potential candidates for reporting.
Conclusions
Underfilled blood tubes, or short draws, may be valid collections for several coagulation, chemistry, and hematology tests—which may prevent additional unnecessary phlebotomy. Laboratories should perform their own studies to determine if short draws are acceptable for limited testing using their tube and instrument types.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Alberta Precision Laboratories, Diagnostic and Scientific Research Centre, Calgary, AB, Canada.
- Departments of Pathology and Laboratory Medicine, Cumming School of Medicine, Health Sciences Centre, Foothills Campus, University of Calgary, Calgary, AB, Canada.
- Department of Cardiac Sciences, Cumming School of Medicine, Health Sciences Centre, Foothills Campus, University of Calgary, Calgary, AB, Canada.
- Department of Laboratory Medicine and Pathology, College of Health Sciences, Faculty of Medicine and Dentistry, University of Alberta, Edmonton, AB, Canada.
ORCID-Profile: Lawrence de Koning
Zitieren
APA
de Koning, L., Ezra, S., Cai, F., Seiden-Long, I., Roshan, T., Gifford, J. L., & Tsui, A. K. Y. (2025). Underfilled tubes revisited: What blood tests can be reported on short draws?. American Journal of Clinical Pathology. https://doi.org/10.1093/ajcp/aqaf109
Vancouver
de Koning L, Ezra S, Cai F, Seiden-Long I, Roshan T, Gifford JL, et al. Underfilled tubes revisited: What blood tests can be reported on short draws?. American Journal of Clinical Pathology. 2025. doi:10.1093/ajcp/aqaf109
BibTeX
@article{deKoning2025,
author = {de Koning, Lawrence and Ezra, Sally and Cai, Fangze and Seiden-Long, Isolde and Roshan, Tariq and Gifford, Jessica L and Tsui, Albert K Y},
title = {Underfilled tubes revisited: What blood tests can be reported on short draws?},
journal = {American Journal of Clinical Pathology},
year = {2025},
doi = {10.1093/ajcp/aqaf109}
}