Technical features and analytical performances of the Stago Expert Preanalytical Check module for hemolysis, icterus, and lipemia measurement
American Journal of Clinical Pathology · 2025
DOI: 10.1093/ajcp/aqaf086 ↗Kurz gefasst
Fachartikel in der Zeitschrift American Journal of Clinical Pathology, 2025. Verfasst von Pascale Laroche und 4 weiteren Autor:innen. Beteiligte Einrichtung: Research and Development Department, Diagnostica Stago, Asnieres sur Seine, France. Hauptthemen laut MEDLINE: Hemolysis, Jaundice, Hyperlipidemias. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Objective
Interference caused by hemolysis, icterus, and lipemia (HIL) are a common source of preanalytical errors during coagulation testing. Stago has developed a specific module on its coagulation analyzers to standardize the detection of HIL.
Methods
The aims of this study were to evaluate the accuracy, carryover, and cross-interference of HIL measurement and compare results with those obtained on cobas 8000 modular analyzer series devices (Roche Diagnostics). These evaluations were carried out using Clinical and Laboratory Standards Institute guidance and regulatory study requirements.
Results
We demonstrated the absence of sample carryover through the HIL measurement chamber and excellent accuracy between HIL theoretical and measured concentrations, with an R2 above 0.99 in all cases. We did not detect any cross-interference of bilirubin and lipids to hemoglobin, of hemoglobin and lipids to bilirubin, or for hemoglobin and bilirubin to lipids. We found excellent agreement between HIL on the Stago STA R Max3 and cobas 8000 platforms, with κ coefficients ranging from 0.928 to 0.963.
Conclusions
These results support the implementation of the Stago Expert Preanalytical Check module for HIL interference measurement in routine coagulation laboratory testing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Research and Development Department, Diagnostica Stago, Asnieres sur Seine, France.
- Medical Affairs Department, Diagnostica Stago, Asnieres sur Seine, France.
- Evaluation Laboratory, Diagnostica Stago, Gennevilliers, France.
Zitieren
APA
Laroche, P., Feriel, J., Tanala Marolahy, G., Chachlica, O., & Depasse, F. (2025). Technical features and analytical performances of the Stago Expert Preanalytical Check module for hemolysis, icterus, and lipemia measurement. American Journal of Clinical Pathology. https://doi.org/10.1093/ajcp/aqaf086
Vancouver
Laroche P, Feriel J, Tanala Marolahy G, Chachlica O, Depasse F. Technical features and analytical performances of the Stago Expert Preanalytical Check module for hemolysis, icterus, and lipemia measurement. American Journal of Clinical Pathology. 2025. doi:10.1093/ajcp/aqaf086
BibTeX
@article{Laroche2025,
author = {Laroche, Pascale and Feriel, Joffrey and Tanala Marolahy, Geneviève and Chachlica, Odile and Depasse, François},
title = {Technical features and analytical performances of the Stago Expert Preanalytical Check module for hemolysis, icterus, and lipemia measurement},
journal = {American Journal of Clinical Pathology},
year = {2025},
doi = {10.1093/ajcp/aqaf086}
}
EUVIMED