Good correlation between tacrolimus concentrations using improved CMIA on the Alinity i analyzer and LC-MS/MS method from a reference laboratory but unexpected negative bias with another LC-MS/MS method from a different reference laboratory

Kelsey Woodard, Tracey Kisler, Amitava Dasgupta

American Journal of Clinical Pathology · 2024

DOI: 10.1093/ajcp/aqae005 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Clinical Pathology, 2024. Verfasst von Kelsey Woodard und 2 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Laboratories, University of Kansas Hospital, Kansas City, KS, US. Hauptthemen laut MEDLINE: Immunosuppressive Agents, Tacrolimus. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 14 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

We compared tacrolimus concentrations obtained by the more recently US Food and Drug Administration–approved tacrolimus assay (CMIA) on the Alinity i analyzer (Abbott Laboratories) with a liquid chromatography/tandem mass spectrometry (LC-MS/MS)–based method from 2 reference laboratories. We also investigated the correlation between the CMIA tacrolimus and Elecsys tacrolimus assays.

Methods

Tacrolimus concentrations were measured in EDTA whole blood by the chemiluminescent microparticle immunoassay (CMIA) using the Alinity i analyzer, and then 2 aliquots were sent to 2 reference laboratories, both using ascomycin as the internal standard for the LC-MS/MS method.

Results

The total precision of the CMIA tacrolimus assay was excellent. When tacrolimus concentrations obtained by the LC-MS/MS method from reference laboratory A were plotted on the x-axis and corresponding CMIA values were plotted on the y-axis, the following regression equation was observed: y = 0.9721x + 1.005 (r = 0.95), indicating no significant bias with the CMIA. However, when tacrolimus values obtained from reference laboratory B were used for comparison, the regression equation was y = 0.7664x + 1.775 (r = 0.93), indicating significant negative bias with the CMIA. When we compared tacrolimus concentrations obtained by reference laboratories A and B, we observed positive bias with tacrolimus concentrations obtained by reference laboratory B. However, tacrolimus concentrations obtained by both CMIA and Elecsys immunoassays were comparable.

Conclusions

Because of good correlation of tacrolimus concentrations using the CMIA and LC-MS/MS from reference laboratory A, our long-term reference laboratory for drug analysis, we concluded that the CMIA on the Alinity i can be used for therapeutic drug monitoring of tacrolimus.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Immunosuppressive AgentsTacrolimusHumansImmunoassayLaboratories, ClinicalLiquid Chromatography-Mass SpectrometryLuminescent MeasurementsTandem Mass Spectrometry

Stichwörter der Autor:innen

LC-MS/MSascomycinbiastacrolimustherapeutic drug monitoring

Substanzen

Immunosuppressive AgentsTacrolimus

Autor:innen und Institutionen

  • Clinical Laboratories, University of Kansas Hospital, Kansas City, KS, US.
  • Department of Pathology and Laboratory Medicine, University of Kansas Medical Center, Kansas City, KS, US.

Zitieren

APA

Woodard, K., Kisler, T., & Dasgupta, A. (2024). Good correlation between tacrolimus concentrations using improved CMIA on the Alinity i analyzer and LC-MS/MS method from a reference laboratory but unexpected negative bias with another LC-MS/MS method from a different reference laboratory. American Journal of Clinical Pathology. https://doi.org/10.1093/ajcp/aqae005

Vancouver

Woodard K, Kisler T, Dasgupta A. Good correlation between tacrolimus concentrations using improved CMIA on the Alinity i analyzer and LC-MS/MS method from a reference laboratory but unexpected negative bias with another LC-MS/MS method from a different reference laboratory. American Journal of Clinical Pathology. 2024. doi:10.1093/ajcp/aqae005

BibTeX
@article{Woodard2024,
  author = {Woodard, Kelsey and Kisler, Tracey and Dasgupta, Amitava},
  title = {Good correlation between tacrolimus concentrations using improved CMIA on the Alinity i analyzer and LC-MS/MS method from a reference laboratory but unexpected negative bias with another LC-MS/MS method from a different reference laboratory},
  journal = {American Journal of Clinical Pathology},
  year = {2024},
  doi = {10.1093/ajcp/aqae005}
}
Kostenlos starten →