The Effects of Inspiratory Flows, Inspiratory Pause, and Suction Catheter on Aerosol Drug Delivery with Vibrating Mesh Nebulizers During Mechanical Ventilation
Journal of Aerosol Medicine and Pulmonary Drug Delivery · 2024
DOI: 10.1089/jamp.2023.0026 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Aerosol Medicine and Pulmonary Drug Delivery, 2024. Verfasst von Hui-Ling Lin und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Respiratory Therapy, Chang Gung University, Taoyuan, Taiwan. Hauptthemen laut MEDLINE: Nebulizers and Vaporizers, Aerosols, Respiration, Artificial, Albuterol. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Beatmung und Weaning
Abstract (Originalsprache)
Background
Some experts recommend specific ventilator settings during nebulization for mechanically ventilated patients, such as inspiratory pause, high inspiratory to expiratory ratio, and so on. However, it is unclear whether those settings improve aerosol delivery. Thus, we aimed to evaluate the impact of ventilator settings on aerosol delivery during mechanical ventilation (MV).
Methods
Salbutamol (5.0 mg/2.5 mL) was nebulized by a vibrating mesh nebulizer (VMN) in an adult MV model. VMN was placed at the inlet of humidifier and 15 cm away from the Y-piece of the inspiratory limb. Eight scenarios with different ventilator settings were compared with endotracheal tube (ETT) connecting 15 cm from the Y-piece, including tidal volumes of 6–8 mL/kg, respiratory rates of 12–20 breaths/min, inspiratory time of 1.0–2.5 seconds, inspiratory pause of 0–0.3 seconds, and bias flow of 3.5 L/min. In-line suction catheter was utilized in two scenarios. Delivered drug distal to the ETT was collected by a filter, and drug was assayed by an ultraviolet spectrophotometry (276 nm).
Results
Compared to the use of inspiratory pause, the inhaled dose without inspiratory pause was either higher or similar across all ventilation settings. Inhaled dose was negatively correlated with inspiratory flow with VMN placed at 15 cm away from the Y-piece ( r s = −0.68, p < 0.001) and at the inlet of humidifier ( r s = −0.83, p < 0.001). The utilization of in-line suction catheter reduced inhaled dose, regardless of the ventilator settings and nebulizer placements.
Conclusions
When VMN was placed at the inlet of humidifier, directly connecting the Y-piece to ETT without a suction catheter improved aerosol delivery. In this configuration, the inhaled dose increased as the inspiratory flow decreased, inspiratory pause had either no or a negative impact on aerosol delivery. The inhaled dose was greater with VMN placed at the inlet of humidifier than 15 cm away the Y-piece.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Respiratory Therapy, Chang Gung University, Taoyuan, Taiwan.
- Department of Respiratory Care, Chang Gung University of Science and Technology, Chiayi, Taiwan.
- Respiratory Care, Department of Cardiopulmonary Sciences, Rush University, Chicago, Illinois, USA.
- Aerogen Pharma Corp, San Mateo, California, USA.
ORCID-Profile: Hui-Ling Lin, James B Fink, Jie Li
Zitieren
APA
Lin, H. L., Fink, J. B., & Li, J. (2024). The Effects of Inspiratory Flows, Inspiratory Pause, and Suction Catheter on Aerosol Drug Delivery with Vibrating Mesh Nebulizers During Mechanical Ventilation. Journal of Aerosol Medicine and Pulmonary Drug Delivery. https://doi.org/10.1089/jamp.2023.0026
Vancouver
Lin HL, Fink JB, Li J. The Effects of Inspiratory Flows, Inspiratory Pause, and Suction Catheter on Aerosol Drug Delivery with Vibrating Mesh Nebulizers During Mechanical Ventilation. Journal of Aerosol Medicine and Pulmonary Drug Delivery. 2024. doi:10.1089/jamp.2023.0026
BibTeX
@article{Lin2024,
author = {Lin, Hui-Ling and Fink, James B and Li, Jie},
title = {The Effects of Inspiratory Flows, Inspiratory Pause, and Suction Catheter on Aerosol Drug Delivery with Vibrating Mesh Nebulizers During Mechanical Ventilation},
journal = {Journal of Aerosol Medicine and Pulmonary Drug Delivery},
year = {2024},
doi = {10.1089/jamp.2023.0026}
}
EUVIMED